background image

19

Maintenance and cleaning

The device is maintenance-free. It is only subject to the natural 
wear and tear of components and their statistical failure rate.

Cleaning:

For cleaning disconnect the main plug!

Use only cleaning agents which have been approved by 

IKA

®

 to 

clean 

IKA

®

 devices.

Dirt

Cleaning agent

Dyes

isopropyl alcohol

Construction materials   water containing tenside/isopropyl alcohol
Cosmetics

water containing tenside/isopropyl alcohol

Foodstuffs

water containing tenside

Fuels

water containing tenside

For materials which are not listed, please request information from 

IKA

®

 

application support.

Wear protective gloves during cleaning the devices.
Electrical devices may not be placed in the cleansing agent for the 
purpose of cleaning.
Do not allow moisture to get into the device when cleaning.
Before using another than the recommended method for clean-
ing or decontamination, you must ascertain with 

IKA

®

 that this 

method does not destroy the device.

Spare parts order:

When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
- item and designation of the spare part see 

www.ika.com

spare parts diagram and spare parts list

- Software version.

Repair:
Please send in device for repair only after it has been 
cleaned and is free from any materials which may consti-
tute a health hazard.

For this, you should request the “

Decontamination Certifi-

cate

” from 

IKA

®

, or use the download printout of it from the 

IKA

®

 website 

www.ika.com

.

If you require servicing, return the device in its original packaging. 
Storage packaging is not sufficient. Please also use suitable trans-
port packaging.

Warranty

In accordance with 

IKA

®

 warranty conditions, the warranty period 

is 24 months. For claims under the warranty please contact your 
local dealer. You may also send the machine direct to our factory, 
enclosing the delivery invoice and giving reasons for the claim. You 
will be liable for freight costs.

The warranty does not cover worn out parts, nor does it apply 
to faults resulting from improper use, insufficient care or mainte-
nance not carried out in accordance with the instructions in this 
operating manual.

Error codes

The fault is shown by an error message in the display (C) as following if the error occurs, e.g. Error 4.
Proceed as follows in such cases:

 Switch the device switch (A) off.

 Remove the stirrer tool and remove the device from the assembly.

 Reduce the speed and switch on (device switch (A)) the device without the stirrer tool.

Error

Cause

Effect

Solution

Error 2

Motor current sensor fault

motor off

- Switch off the device

Error 3

Temperature inside device is too high

motor off

- Switch off the device and allow it to cool down

Error 4

Motor blockage or overload

motor off

- Switch off the device
- Decrease the load on the motor and restart again

Error 8

Speed sensor fault or overload

motor off

- Switch off the device

Error 21

Safety relay fault

motor off

- Switch off the device

If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps: 
- Contact the service department 
- Send the device for repair, including a short description of the fault. 

Summary of Contents for EUROSTAR 100 P4 control

Page 1: ...20000015715 IKA EUROSTAR 100 P4 control EUROSTAR 100 P4 control_112017 Operating instructions EN...

Page 2: ...4 Fig 2 Fig 1 A J I H G F E EUROSTAR 100 P4 control station Wireless Controller WiCo Q B M O N C L K P D Fig 3 H I J G G Fig 4 K L M Fig 5 T U Q S U Fig 7 X I G Z H G W V Fig 8 Fig 6 RS 232 USB Temp...

Page 3: ...Note for Canada IC This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 This device may not cause interference 2 This device...

Page 4: ...y with design requirements on the part of the product For this reason it may become necessary for users to take other precautionary safety measures For example glass device or other stirring vessels t...

Page 5: ...WiCo is fully discharged the only means of switching the station off are the safe STOP and the off switch Please note the following safety in structions for the battery pack RB 1 re chargeable batter...

Page 6: ...ufacturer If the device is operated improperly or contrary to the manufac ture s specifications if the device or the printed circuit board are modified by third parties Drive The rotary knob B see Fig...

Page 7: ...that it is not liable to topple and does not start to move during the stirring procedure Securing The clamping chuck and output shaft permit clamping IKA recom mend stirrer tools see section Permitte...

Page 8: ...OFF and ON If the Bluetooth function of the WiCo is active you can use the Bluetooth search button H see Fig 1 to search for the WiCo A beep signal will be audible even if the WiCo is switched off Th...

Page 9: ...xplanation of symbols on the working screen The symbols displayed change depending on the status and settings of the WiCo The screen be low shows the most significant symbols on the working screen The...

Page 10: ...in the display Torque Calibration Stirring Intermittent Mode Torque Display Torque Limit Speed Limit Stirring Speed Limit Speed Limit 300 rpm OK Menu Back Menu Back Back B 0 0 actual rpm set rpm Torq...

Page 11: ...top CW CCW Interval English Deutsch Temperature Background Brightness Firmware Update Info Volume Key Tone Device name Remote Control Bluetooth Stirring Safety Software Firmware Update Info 300 rpm 32...

Page 12: ...tives Torque Display The Torque Display menu allows you to specify that the torque is shown in the display A tick shows that the option is activated Note Pressing the Back button while the device is r...

Page 13: ...ed In addition you could define the intermittent mode is activated or not in the programs If the intermittent mode is activated the run time stop time value from Intermittent Mode setting will be take...

Page 14: ...will be shown rpm 150 300 Seg No 1 rpm 150 Time hh mm 00 10 Interm Mode Yes PROGRAM 1 Detail rpm 150 300 PROGRAM 1 Detail rpm 150 300 Seg No 2 rpm 150 Time hh mm 00 10 Interm Mode No PROGRAM 1 Detail...

Page 15: ...Interm Mode Edit Delete Insert 1 2 3 4 5 100 0 00 10 00 00 Yes No Programs Start Edit Delete Program 1 Program 2 Program 3 Program 4 Program 5 PROGRAM 1 Seg No rpm Time hh mm Interm Mode 1 2 3 4 5 0 0...

Page 16: ...he Languages option allows you to select the desired lan guage by turning and pressing the rotary push knob B A tick indicates the language that is set for the system Units The Units option allows you...

Page 17: ...ce only at the computer s request Command syntax and format The following applies to the command set Commands are generally sent from the computer Master to the stirrer machine Slave The stirrer machi...

Page 18: ...station to the computer A USB micro B USB A 9 pin RS 232 9 pin RS 232 or micro B A USB micro B USB A micro B or A 9 pin RS 232 9 pin RS 232 USB micro B USB A micro B USB A USB A USB A USB 2 0 cable m...

Page 19: ...require servicing return the device in its original packaging Storage packaging is not sufficient Please also use suitable trans port packaging Warranty In accordance with IKA warranty conditions the...

Page 20: ...000 R 1331 Anchor stirrer 1000 R 1333 Anchor stirrer 800 R 1375 Paddle stirrer 800 R 1376 Paddle stirrer 800 R 1342 Propeller stirrer 2000 R 1345 Propeller stirrer 800 R 1381 Propeller stirrer 2000 R...