Error codes
53
Err. 13 – Hotplate safety sensor, open-circuit
Causes
› target / current different of the adjustable safety circuit
for minimum temperature monitoring
Effect
› heating switched off
Solutions
› after switching on, change the SAFE TEMP to a different
value; if this solves the issue, the previous value can be
reset by switching the device off and on again
Err. 14 – External temperature sensor, short-circuit
Causes
› short circuit in temperature sensor plug
› short circuit in the cable or temperature sensor
Effect
› heating switched off
Solutions
› check the plug
› replace the temperature sensor
Err. 21 – Fault during heating plate safety test
Causes
› safety relay does not open
Effect
› heating switched off
Solutions
› switch device off, leave to cool and switch on again
Err. 22 – Fault during heating plate safety test
Causes
› S_CHECK cannot generate H_S_TEMP
Effect
› heating switched off
Solutions
› switch device off, leave to cool and switch on again
Err. 24 – Heating plate temperature
is higher than the set safety temperature
Causes
› SAFE TEMP H (Hotplate) has been set to lower than the
current temperature of the heating plate
› disconnection of heating plate control temperature sensor
Effect
› heating switched off
Solutions
› leave the heating plate to cool
› set the SAFE TEMP H (Hotplate) higher
Summary of Contents for C-MAG HS7
Page 2: ......
Page 3: ...58 4 166 112 ESPA OL ENGLISH FRAN AIS PORTUGU S...
Page 4: ......
Page 5: ...ENGLISH...
Page 10: ......
Page 21: ......
Page 29: ...Menu 29 05 00 00 00 mm ss 00 03 mm ss Clockwise continuous Factory settings...
Page 31: ...Menu 31 SAFE Factory settings...
Page 58: ......
Page 59: ...ESPA OL...
Page 64: ......
Page 75: ......
Page 83: ...Men 83 Configuraci n de f brica 05 00 00 00 mm ss 00 03 mm ss hacia la derecha continuo...
Page 85: ...Men 85 Configuraci n de f brica SAFE...
Page 112: ......
Page 113: ...PORTUGU S...
Page 118: ......
Page 129: ......
Page 137: ...Menu 137 Configura o de f brica 05 00 00 00 mm ss 00 03 mm ss em sentido hor rio cont nuo...
Page 139: ...Menu 139 SAFE Configura o de f brica...
Page 166: ......
Page 167: ...FRAN AIS...
Page 172: ......
Page 183: ......
Page 191: ...Menu 191 R glage d usine 05 00 00 00 mm ss 00 03 mm ss Dans le sens horaire continu...
Page 193: ...Menu 193 R glage d usine SAFE...
Page 220: ......