52
Error codes
Err. 3 – Temperature inside unit is higher than + 80 °C
Causes
› heat accumulation between heating plate and housing
› permitted ambient temperature exceeded
Effect
› heating switched off
Solutions
› switch device off, leave to cool and switch on again
› change experiment
› observe maximum permissible ambient temperature
Err. 4 – Motor control is unavailable
Causes
› motor blocked or overloaded
Effect
› heating switched off
› motor switched off
Solutions
› reduce load torque or use smaller magnetic rods
› reduce target speed
Err. 5 – No temperature increase measured by temperature sensor
(selected time in menu)
Causes
› sensor not in medium
› volume of medium to be measured too large
› heat conductivity of medium to be measured too low
› heat conductivity of the vessel is too low
› in the case of indirect heating, the overall heat
conductivity resistance is too large
Effect
› heating switched off
Solutions
› place the sensor in the medium
› reduce the volume of the media
› use a carrier fluid with better heat conductivity properties
› replace the glass vessel with a metal pot
› increase the “Time-out” period
Err. 6 – Sensor plug out during Heating
Causes
› defective connection cable
Effect
› heating switched off
Solutions
› replace cable
Summary of Contents for C-MAG HS7
Page 2: ......
Page 3: ...58 4 166 112 ESPA OL ENGLISH FRAN AIS PORTUGU S...
Page 4: ......
Page 5: ...ENGLISH...
Page 10: ......
Page 21: ......
Page 29: ...Menu 29 05 00 00 00 mm ss 00 03 mm ss Clockwise continuous Factory settings...
Page 31: ...Menu 31 SAFE Factory settings...
Page 58: ......
Page 59: ...ESPA OL...
Page 64: ......
Page 75: ......
Page 83: ...Men 83 Configuraci n de f brica 05 00 00 00 mm ss 00 03 mm ss hacia la derecha continuo...
Page 85: ...Men 85 Configuraci n de f brica SAFE...
Page 112: ......
Page 113: ...PORTUGU S...
Page 118: ......
Page 129: ......
Page 137: ...Menu 137 Configura o de f brica 05 00 00 00 mm ss 00 03 mm ss em sentido hor rio cont nuo...
Page 139: ...Menu 139 SAFE Configura o de f brica...
Page 166: ......
Page 167: ...FRAN AIS...
Page 172: ......
Page 183: ......
Page 191: ...Menu 191 R glage d usine 05 00 00 00 mm ss 00 03 mm ss Dans le sens horaire continu...
Page 193: ...Menu 193 R glage d usine SAFE...
Page 220: ......