Commis-
sioning
Put device switch (A) in the OFF position
Plug in (G) mains cable
Once connected to the power supply the device is in
“stand-by” mode
Put device switch (A) in the ON position
Set the engine speed with the operating button (MOT)
Any set values are retained when device is switched
off and even after the device is disconnected from
the power!
Commissioning
Stirring
Heating
without
external
temperature sensor
not in operating
mode A
with
external
temperature sensor
(direct
temperature
control in the
medium)
Put device switch (A) in the OFF position
Plug in the PT 1000 sensor into the socket (I)
Put device switch (A) in the ON position
Adjust the set-point temperature with the operating
button (TEMP)
• LED (F) lights up
• the instantaneous temperature (K) is indicated on the dis-
play (corresponds sensor- and/or medium temperature)
• indicated set point temperature (L) corresponds the
desired medium temperature
• LED (E) lights up and the symbol for energy input (M) is
indicated on the display (with the heating is switched on)
• indicate HOT (D) on the display with over 50 °C on the
heating plate (in the agitating and stand-by-mode)
Put device switch (A) in the ON position
Set the operating mode (see capture “Operating modes”)
Adjust the set-point temperature with the operating
button (TEMP) (only be set in operating mode B)
• indicate the adjusted set point temperature on the
LCD display (D) position (L)
• indicated the set point temperature corresponds the
heating plate temperature
• LED (E) lights up and the symbol for energy input
(M) is indicated on the display (with the heating is
switched on)
•
indicate HOT (D) on the display with over 50 °C on
the heating plate (in the agitating and stand-by-mode)
Any malfunctions during operation will be identified by an error message
on the display (D).
Proceed as follows in such cases:
•
Switch off device using the main switch (A)
•
Carry out corrective measures
•
Restart device
Error codes
Error Cause
Solution
code
E1
Sensor not
• Connect sensor
connected
• Change operating mode
(operating mode A)
E3
Temperature inside • Switch off device and allow
device too high
to cool down
E5
Sensor not in
• Sensor in medium
media
immerse
E6
Sensor defectively
• Exchange the sensor
E7
Hi-Temp
• Medium to cool down
temperature border
exceeded
E9
Initializations error
• Switch off/on the device
E11
Motor blocks
• use other stirring bar
• reduce viscosity
If the actions described fail to resolve the fault or another error code is
displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
Summary of Contents for C-MAG HS 10
Page 2: ...2 E D A B C H G I F Fig 1 J K M L N...
Page 29: ...29 4 29 30 30 31 31 HI TEMP 31 31 32 32 33 33 33 34 500 0 500 C 25 C IKA RU...
Page 30: ...30 PT 1000 20 PT 1000 2 3 300 400 19 800 PT 1000...
Page 41: ...41 41 42 42 43 43 HI TEMP 43 43 44 44 45 45 45 46 500 0 500 25 JA...
Page 47: ...47 47 48 48 49 49 HI TEMP 49 49 50 50 51 51 51 52 500 0 500 25 IKA KO...
Page 76: ...76 h 0 500 C 25 IKA PT 1000 ETS D 20 PT 1000 B C PTFE 300 400 C PTFE 2 3 19 a a a a a...