27
”Evaluation“ tab
Default evaluation:
After a measurement has been completed,
as a rule the results are extended and corrected in accordance
with the applicable national or international standards.
Select from the list:
The selected standard will be applied to all the subsequent mea-
surements.
Evaluated measurements will comply with this standard: if a mea-
surement has already been evaluated, the evaluation procedure
will be used for any subsequent re-evaluation. Otherwise the cur-
rent setting for the evaluation procedure will be used.
Measurements evaluated previously are not affected by this
change.
Display unit:
This sets the display units for the calorimetric re-
sults. Select them from the list.
The selected range of units applies to the display of energy values
[J], calorific values [J/g] and thermal capacities [J/K].
Reference combustion value:
Input here the calorific value of
the substance used for calibration. As a rule this substance is ben-
zoic acid with a calorific value certified by the manufacturer.
Evaluation
Other settings