Connection of peripherals
RS 232 PC
Serial connection for controlling the calorimeter using CalWin
®
or
operatingwith serial printer C 1.50.
Standard setting: Baud rate:
9600
Data
bits:
8
Stop
bits:
1
Parity:
none
Handshake:
none
RS 232 BALANCE
Interface for connecting scales (Mettler, Ohaus, Sartorius, Kern).
Standard setting: Baud rate:
1200
Data
bits:
7
Parity:
odd
Stop
bits:
1
Handshake:
none
ETHERNET
Network connection for data transfer via the network, e.g. to a net-
work printer.
USB-DEVICE
USB interface (only for service).
USB-HOST
Interface for connection of a printer or with a USB mouse.
On/off switch
NOTICE
Switch the device off at the on/off
switch only when directed to do so
by the menu command.
Data may be lost if the appliance is
switched off other than by follow-
ing the menu commands.
The device is switched on and off via the on/off switch located on
the right-hand side.
Switch the device off at the on/off switch.
Lift moves upwards.
Switch the appliance off only by using the menu commands.
Lift moves downwards.
NOTICE
While the peripheral devices are
being connected, they and the
calorimeter must be turned-off on
the mains power switch!