23
Factory Settings:
Select the ”Factory Settings” option by turning and pres-
sing the rotary/push knob. The system requests confirmati-
on to recreate the factory settings. Pressing the ”OK” but-
ton resets all the system settings to the original standard
values set at dispatch from the factory (see chapter „13.5
Menu structure“).
Communication:
In the “Settings: communication, device name” menu, you
can enter individual names for the devices. The names ap-
pear on the start screen after switching on the device.
Device
Timer
Functions in interval mode
Stirrer
yes
start/stop oder CW/CCW
LED panel
yes
on/off
Temperature
-
-
pH sensor
-
-
Pumps
yes
start/stop,
Gas
yes
open/close
General timer functions
Activate timer:
The set time starts as soon as you exit the
menu.
After the time has expired the functions that
were running also stop.
Interval mode:
CW
Clockwise
CCW
Counterclockwise
Time 1 Indicates how long the status of function 1 is switched
on and function 2 is switched off.
Time 2 Indicates how long the status of function 2 is switched
on and function 1 is switched off.
The set time 1 with corresponding function starts as soon as you
press the “Start” button.
After time 1 has expired, the function stops and time 2 begins to
run with the corresponding function.
After time 2 has expired, time 1 starts again with the correspond-
ing function.
If a timer and interval mode is set at the same time, the interval
mode ends when the timer has expired. If interval mode is only
activated, without the timer being active, then interval mode must
Timer
The timer functions and the interval modes are available in the
following menus:
be ended manually.
Interval mode example for stirrer:
Symbol rotation
direction
Chuck rotation
direction
Graph
CW
Time
Speed
Factory setting: continuous mode
CW
Time
Speed
Function “run/stop“ activated:
:
The run time and stop time can be set separately.
CW/CCW
Time
Speed
Function “CW/CCW“ activated:
The clockwise and counterclockwise running
time can be set independently.
∞
/ CW...
Run-Stop-Run / CW ...
Run (CW) / Run (CCW) ...
When using multiple devices with different settings, giving
the devices a name can be useful. It helps with identifica-
tion and communication, for example, with a PC via USB
interfaces, RS 232 and Bluetooth (see chapter “15 Inter-
faces and outputs”).
Informationen:
The ”Information” option offers the user an overview of
the most important system settings of the Algaemaster 10.