6
072010
H44
Cross sleeve
ETS-D5
Contact thermometer
Installing the support rod
The support rod is attached using the thraeded support bore (F).
• Screw nut M10 on to the support rod as far
as the stop
• Screw on the support rod as far as the
stop by hand
• Tighten the support rod and nut M10
using a flat wrench (SW17).
• Use bossheads to assemble
accessory parts or acces-
sory devices.
The appliance is maintenance-free.
Cleaning
Only use cleansing agents which have been recommended by
IKA
.
Use to remove:
Dyes
isopropyl alcohol
Construction materials water containing tenside/ isopropyl alcohol
Cosmetics
water containing tenside/ isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
Footstep during the enterprise a disturbance up, is indicated these
with the devices to
HS
and
HP
by an error message in the display
(C). Whenever an error message appears switch the device off to
cool down.
Error codes
Error
Cause
Solution
code
E1
Inner temperature
• switch off the unit and allow it
too high
cool down
E6
Motor seized
• Stirring bars ungeeignet
• Reduce the viscosity
• Foreign bodies in the equipment
remove
E9
Safety circuit
• Plug the contact plug (G)
ETC
• Plug the contact thermometer
• Connecting cable, plug or
contact thermometer defect,
exchange
Have the device repaired if the error is not corrected using the
measures described or if another error code is displayed.
Stirring organs
stirring bars:
Ø 6 mm, lengts to 15mm
Ø 8 mm, lengts to 50mm
Ø10mm, lengt to 80mm
Any other accessories
RS 1
Set of magnetic stirring bars
RSE
Stirring bar remover
H16V
Support rod
H16.1
Extension
H38
Holding rod
Accessories
Maintenance
Fig. 3