background image

28

RF emissions

CISPR 11

Group 1

The device uses RF energy 

only for its internal function. 

Therefore, its RF emissions are 

very low and are not likely to 

cause any interference in nearby 

electronic equipment.

RF emissions

CISPR 11

Class B

The device is suitable for use in 

all establishments, other than 

domestic and those directly con-

nected to the public low-voltage 

power supply network that sup-

plies buildings used for domestic 

purposes.

Harmonic emissions IEC 

61000-3-2

Class A

Voltage fluctuations/

flicker emissions

IEC 61000-3-3

Complies

Table 2

Guidance and manufacturer’s declaration – electromagnetic immunity

The device is intended for use in the electromagnetic environment specified below. 

The customer or the user of the device should assure that it is used in such an 

environment.

Immunity test

IEC 60601 test 

level

Compliance 

level

Electromagnetic 

environment - 

guidance

Electrostatic dis-

charge (ESD)

IEC 61000-4-2

±8 kV contact

±15 kV air

±8 kV contact

±15 kV air

Floors should be 

wood, concrete or 

ceramic tile. If floors 

are covered with 

synthetic material, 

the relative humidity 

should be at least 

30%.

Electrical fast 

transient/burst 

IEC 61000-4-4

power supply 

lines: ±2 kV

input/output 

lines: ±1 kV

power supply 

lines: ±2 kV

Mains power quality 

should be

that of a typical 

commercial or

hospital environ-

ment.

Summary of Contents for iMove

Page 1: ...automatic BLOOD PRESSURE MONITOR model iMove model iMove...

Page 2: ...ur blood pressure and pulse correctly For all questions con cerning the device please contact your local distributor or TM iHerz service centre in your country ATTENTION This blood pressure monitor is...

Page 3: ...one which corresponds to the heartbeat pushing blood into the arteriae and a diastolic lower one which means the blood pressure between two heartbeats Blood pressure is subject to fluctuations during...

Page 4: ...Organization WHO Category Systolic Blood Pressure mmHg Diastolic Blood Pressure mmHg Blood pressure too low 100 60 Blood pressure optimum 100 119 60 79 Blood pressure normal 120 129 80 84 Blood press...

Page 5: ...to 6 ranges described in the table of the section Which values are normal After the measurement there appears the dotted line in the left part of the display opposite the zone to which the result of t...

Page 6: ...children The device is not intended for patient transport outside a healthcare facility The device is not intended for public use This device is intended for no invasive measuring and monitoring of a...

Page 7: ...more than 3 minutes applied to the wrist may lead to an ecchymosis Please check that operation of the device does not result in prolonged impairment of patient blood circulation When measurement pleas...

Page 8: ...ent If you are allergic to polyester nylon or plastic please don t use this device During use the patient will be in contact with the cuff The materials of the cuff have been tested and found to compl...

Page 9: ...uals at authorized sales service centers Please report to Transtek if any unexpected operation or events occur Keep the unit out of reach of infants young children or pets to avoid inhalation or swall...

Page 10: ...herwise the performance and lifetime of the device will be impacted and reduced KEEP THIS MANUAL DURING THE WHOLE LIFETIME OF THE DEVICE COMPONENTS OF YOUR BLOOD PRESSURE MONITOR KIT The blood pressur...

Page 11: ...lay does not light up Do not use new and used batteries together Do not use different types of batteries together Do not dispose the batteries in fire Batteries may explode or leak Remove batteries if...

Page 12: ...which group of memory it is kPa Measurement Unit of the blood pressure mmHg Measurement Unit of the blood pressure Low battery Batteries are low and need to be replaced Irregular heartbeat Blood pres...

Page 13: ...ach record that is stored in the memory The setting range of the year 2019 2057 time for mat 12 H 24H 1 When the monitor is off hold pressing MEM button for about 3 seconds to enter into setting mode...

Page 14: ...atrium of the heart Efforts by the patient to support the wrist can increase the blood pressure Make sure you are in a comfortable relaxed position and do not activate any muscles in the wrist during...

Page 15: ...tested wrist resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths 7 Patien...

Page 16: ...whole measurement Take User 1 for example Press the START STOP to power off otherwise it will turn off within 1 minute MEMORY FUNCTION Reading Measurement Results 1 When the monitor is off press MEM b...

Page 17: ...it e g 2 becomes 3 and so on and the last record 60 is dropped Delete Memories If you did not get the correct measurement you can delete all results by following steps below 1 In the memory mode hold...

Page 18: ...r wrist blood pres sure monitor If the products not operating as you think it should check here before arranging for service PROBLEM SYMPTOM CHECK THIS REMEDY No power Display is dim or will not light...

Page 19: ...nt can affect the measurement Relax for a moment and then measure again E 20 shows The measure ment process does not detect the pulse signal Loosen the clothing on the wrist and then measure again E 2...

Page 20: ...ge SYS 60mmHg to 230mmHg or DIA 40mmHg to 130mmHg or Pulse 40 199 pulse minute CARE AND MAINTENANCE In order to get the best performance please follow the instructions below Put in a dry place and avo...

Page 21: ...tact with the patient s skin with the use of a soft cloth soaked in ethyl alcohol 75 90 and wait for the pressure cuff to dry up Cleaning of the pressure cuff after 200 uses is recommended Periodic ca...

Page 22: ...rrent Caution consult accompanying documents Dispose of in accordance with the requirements of your country The cuff is type BF applied part European representative CE mark Recycle The Green Dot is th...

Page 23: ...4kPa Pulse value 5 Normal working condition A temperature range of 5 C to 40 C A relative humidity range of 15 to 90 non condensing but not requiring a water vapour partial pressure greater than 50 hP...

Page 24: ...ARNING No modification of this equipment is allowed Conformance to standards DSTU EN 1060 1 2015 Non Invasive Sphygmomanometer General Requirements Paragraph 9 2 DSTU EN 1060 3 2015 Non Invasive Sphyg...

Page 25: ...A1 2013 IEC 60601 1 2005 A1 2012 Medical electrical equip ment Part 1 General requirements for basic safety and essential performance EN 60601 1 11 2015 IEC 60601 1 11 2015 Medical electrical equipme...

Page 26: ...tiga tion EN 1060 4 2004 Non invasive sphyg momanometers Part 4 Test procedures to determine the overall system accuracy of automated non invasive sphygmoma nometers ISO 81060 2 2013 Non invasive sphy...

Page 27: ...not use a mobile phone or other devices that emit electro magnetic fields near the unit This may result in incorrect opera tion of the unit 3 Caution This unit has been thoroughly tested and inspecte...

Page 28: ...dance and manufacturer s declaration electromagnetic immunity The device is intended for use in the electromagnetic environment specified below The customer or the user of the device should assure tha...

Page 29: ...hase at 0 0 UT 300 cycle Mains power quality should be that of a typical commercial or hospital environ ment Power frequency 50Hz 60Hz magnetic field IEC 61000 4 8 30 A m 50Hz 60Hz 30 A m 50Hz 60Hz Po...

Page 30: ...kHz 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 7 GHz d 2 3 P where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer d is the...

Page 31: ...the frequency range 150 kHz to 80 MHz field strengths should be less than 3V m Table 4 Recommended separation distances between portable and mobile RF communications equipment and the device The devi...

Page 32: ...ding to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic...

Page 33: ...FRS 460 FM c 5kHz deviation 1kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modula tion b 217Hz 0 2 0 3 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modula tion...

Page 34: ...to FM modulation 50 pulse modulation at 18 Hz may be used because while it does not represent actual modulation it would be worst case The MANUFACTURER should consider reducing the minimum separation...

Reviews: