34
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance be
-
tween the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may
be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may
be used because while it does not represent actual modulation, it would be
worst case.
The MANUFACTURER should consider reducing the minimum separation
distance, based on RISK MANAGEMENT, and using higher IMMUNITY TEST
LEVELS that are appropriate for the reduced minimum separation distance.
Minimum separation distances for higher IMMUNITY TEST LEVELS shall be
calculated using the following equation:
E=
Where P is the maximum power in W, d is the minimum separation distance in
m, and E is the IMMUNITY TEST LEVEL in V/m.