Problem
LCD shows low battery
symbol
LCD shows “Er 0”
LCD shows “Er 1”
LCD shows “Er 2”
LCD shows “Er 3”
LCD shows “Er 4”
LCD shows “Er 5”
LCD shows “Er 6”
LCD shows “Er 7”
LCD shows “Er 8”
LCD shows “Er A”
No response when you press
button or install batterypiles
Low Battery
Pressure system is unstable before
measuremente
Pneumatic system blocked or cuff is too
tight during inflation
More than 3 minutes with cuff pressure
above 15 mmHg
Incorrect operation or strong
electromagnetic interference.
EEPROM accessing error
Pneumatic system leakage or cuff is too
loose during inflation
Fail to detect systolic pressure
Cuff pressure above 300mmHg
Device parameter checking error
Pressure sensor parameter erroron
Fail to detect diastolic pressure
Change all the batteries
Try again without moving
Take out the batteries, wait
five minutes and then
reinstall batteries
Apply the cuff correctly
and try again
Measure again after 5
minutes. If the error is still
there, please contact your
local iHealth distributor
Explanation
Solution
9
Summary of Contents for Track
Page 65: ......
Page 80: ...PT Monitor de Tens o Arterial Autom tico de Bra o MANUAL DO UTILIZADOR...
Page 95: ...Volledig Automatische Online Armbloeddrukmeter GEBRUIKERSHANDLEIDING NL...
Page 110: ...www ihealthlabs eu support GR...
Page 111: ...1 2 3 4 5 6 7 8 9 iHealth 10 Bluetooth...
Page 112: ...iHealth MyVitals iHealth MyVitals App Store Google Play iHealth...
Page 114: ...2 cm 200 3 1 2 1 2 cm...
Page 115: ...22 42 cm 4 iHealth MyVitals smartphone 5 i START STOP ii iii iv iHealth MyVitals...
Page 118: ...Er 0 Er 1 Er 2 Er 3 Er 4 Er 5 Er 6 Er 7 Er 8 Er A iHealth 300 mmHg 15 mmHg...
Page 119: ...www ihealthlabs eu 9 1 2 3 4 25 0 14 s 53 5 6 Hi Lo...
Page 120: ...IEC 60601 1 8 7 10 1 2 3 4 5 6...
Page 121: ...7 8 9 75 90 10 10 000 1 000 11 iHealth www ihealthlabs eu email support i healthlabs eu...