background image

Temperature sensor

TM45x1

5

Welding operation:

Fig. 1: Applying the fixing points

Fig. 2: Applying the welding seams

u

Apply the fixing points of the welding thermowell at equal distance (crosswise) with a sufficient
adhesive force.

u

Apply the welding seams between the fixing points opposite each other.

u

Ensure sufficient intervals between the individual welding operations to avoid glowing through or
warping of the welding thermowell due to overheating.

After completion of the welding operation:

u

Let the workpiece cool down.

u

Slightly grease the thread using a lubricating paste which is suitable and approved for the
application.

u

Screw the sensor into the adapter and tighten. Tightening torque: 30...50 Nm.

5 Electrical connection

The unit must be connected by a qualified electrician. Voltage supply according to EN50178,
SELV, PELV.

u

Disconnect power.

u

Connect the unit as follows:

4

3

2

1

1
2
3
4

Fig. 3: Wiring diagram (colours in accordance with DIN EN 60947-5-2)

1:

BN (brown)

2:

WH (white)

3:

BU (blue)

4:

BK (black)

6 Maintenance, repair and disposal

The operation of the unit is maintenance-free.

Only the manufacturer is allowed to repair the unit.

u

After use dispose of the device in an environmentally friendly way in accordance with the
applicable national regulations.

u

Define regular calibration intervals according to the process requirements.