40
10.5.10 Time-delay empty pipe detection
►
Select [dEP�E] and set the delay time from 0…30 s, at which the signal
should be provided when the pipe is empty�
►
Select [dEP�F] and set the delay time from 0…30 s, at which the signal
should be provided when the pipe is full�
Menu EPD:
[dEP�E]
[dEP�F]
10.5.11 Empty pipe detection limit value
►
Select [EP�Pr] to display the current value of the empty pipe detection in
percent�
►
Select [EP�SP] and set the switch point of empty pipe detection�
Menu EPD:
[EP�Pr]
[EP�SP]
10.5.12 Counting method of the totaliser
►
Select [FPro] and set the value:
–
+ = totalling the volumetric flow values with the correct sign�
0+ = totalling only positive volumetric flow values�
Menu CFG:
[FPro]
10.5.13 Low flow cut-off
►
Select [LFC] and set the limit value�
Menu CFG:
[LFC]
10.6 Service functions
10.6.1 Read min/max values
►
Select [HI�x] or [LO�x] and read the value�
HI�F = maximum volumetric flow, LO�F = minimum volumetric flow
HI�T = maximum temperature, LO�T = minimum temperature
Delete memory:
►
Select [HI�x] or [LO�x]�
►
Briefly press [Enter]�
► keep [▲] or [▼] pressed until [----] is displayed.
►
Briefly press [Enter]�
It makes sense to delete the memories as soon as the unit operates
under normal operating conditions for the first time�
Menu MEM:
[HI�F]
[LO�F]
[HI�T]
[LO�T]