5 Installation
►
Before installing and removing the unit make sure that no pressure is
applied to the system and there is no medium in the pipe�
►
Note dangers related to extreme machine / medium temperatures�
►
In case of a recessed sensor installation, please use an open ring
spanner or a hexagon socket with the corresponding inside contour�
►
During sensor installation, do not exert any axial leverage with the
tool (1) on the rotatable filter cover (2)
(→ Fig. 1)
�
►
In case of installation by means of a hexagon socket (3), make sure
that the tool's spanner flat has an inner height of at least 30 mm
(→ Fig. 2)
�
Align the rotatable filter cover so that it is flush with one of the spanner
flats�
≥ 30mm
1
2
►
Insert the unit in a process connection�
►
Tighten firmly� Recommended tightening torque:
Process connection
Tightening torque in Nm
G1A Aseptoflex Vario
35
G1A sealing cone
20
Depends on the sealing type, the pressure load and the lubrication!
A zero-point calibration of the measured value is possible via the teach