6.2 Installation of the probe
The rod is not included in the scope of supply. It has to be ordered separately (→
3 Items supplied)�
Fixing of the rod:
►
Screw the rod to the unit and tighten it�
Recommended tightening torque:
4 Nm�
For ease of installation and removal the rod
connection can be rotated without restriction�
Even if rotated several times there is no risk
of damage to the unit�
6
In case of high mechanical stress (strong vibration, moving viscous media) it
may be necessary to secure the screw connection, e�g� by a screw retaining
compound�
Substances such as screw retaining compounds may migrate into the
medium� Make sure that they are harmless�
When using mechanical means of securing (e�g� tooth lock washer), protruding
edges must be avoided� They may cause interference reflection�