background image

5.2  Mounting

The unit is mounted using a mounting clamp or an adapter� The following standard 

mounting accessories are available:

Order no� Material

E43000

mounting clamp (for non-pressurised tanks) 

E43001

flange plate 73-90 with clamp adapter (pressure rating of 0�5 bar)

E43002

welding adapter with clamp adapter (pressure rating of 0�5 bar)

E43003

mounting adapter G¾ with clamp adapter (stainless steel, pressure rating of 0�5 bar)

E43004

mounting adapter G1 with clamp adapter (stainless steel, pressure rating of 0�5 bar)

E43005

 flange plate 100-125 with clamp adapter (aluminium, pressure rating of 0�5 bar)

E43006

flange plate 65-80 with clamp adapter (aluminium, pressure rating of 0�5 bar)

E43007

flange plate 54 - 52x52 with clamp adapter (aluminium, pressure rating of 0�5 bar)

E43016

mounting clamp with mounting bracket (for non-pressurised tanks)

E43019

mounting adapter G¾ with clamp adapter (brass, pressure rating of 0�5 bar)

Further accessories on request�

5.3  Set the installation depth

MET

L3

A

H

L1

L

L2

S

M

A = response level
H = tank height
L = probe length
L1 = installation depth
L2 = extension length
L3 = installation depth (min�: 60 mm)
M = height mounting adapter
MET = max� immersion depth
(Active zone) = 31 mm
S = height of the connection piece

 

► Select the installation depth L1 so that the MET zone is covered when the 

response level A has been reached, that is:  

L1 = (H - A) + MET + S�

Summary of Contents for efector160 LI214 Series

Page 1: ...Operating instructions Binary level sensor LI214x 704776 00 10 2010 UK...

Page 2: ...rinciple 4 4 2 Features of the unit 5 5 Installation 5 5 1 Installation location environment 5 5 2 Mounting 6 5 3 Set the installation depth 6 6 Electrical connection 7 7 Operating and display element...

Page 3: ...t the product is suitable for your application without any restrictions If the operating instructions or the technical data are not adhered to personal injury and or damage to property can occur Insta...

Page 4: ...te mounting accessories chapter 5 2 Installation procedure 4 Function 4 1 Measuring principle The unit operates to the capacitive measuring principle By direct contact with the medium the unit detects...

Page 5: ...le detection of media with a very low dielectric constant e g oil The installation depth determines the level 5 3 The unit is virtually immune to almost any kind of build up on the probe The position...

Page 6: ...ssure rating of 0 5 bar E43006 flange plate 65 80 with clamp adapter aluminium pressure rating of 0 5 bar E43007 flange plate 54 52x52 with clamp adapter aluminium pressure rating of 0 5 bar E43016 mo...

Page 7: ...age supply to EN 50178 SELV PELV Disconnect power Connect the unit as follows 3 1 4 2 L L 1 4 3 Pin1 L Pin4 switching output NC Pin3 L Core colours of ifm sockets 1 BN brown 3 BU blue 4 BK black 7 Ope...

Page 8: ...rocess the yellow LED is lit con tinuously unit is in the operating mode sensor zone not covered 8 2 Set to the full vessel The empty adjustment is sufficient for media with a low dielectric constant...

Page 9: ...ation When the supply voltage has been applied the unit is in the operating mode It carries out its measurement and evaluation functions and switches the output Operation indication by LED LED Status...

Page 10: ...nstant medium 1 8 Max tank pressure bar 0 51 Housing materials PP GF orange TPE U Materials wetted parts PP Protection IP 65 IP 67 Protection class II Ambient temperature C 0 60 Medium temperature oil...