Guidelines and manufacturer’s declaration – electromagnetic immunity
The blood pressure measuring device Mobil-O-Graph
®
is intended for use in the environment specified
below. The customer or user of the blood pressure measuring device Mobil-O-Graph
®
should ensure that it is
used in such an environment.
Immunity test
IEC 60601-1 test level
Compliance level
Electromagnetic environment
– guidelines
Electrostatic
discharge (ESD)
according to IEC
61000-4-2
+- 8 kV contact discharge
+- 15 kV air discharge
+- 8 kV contact discharge
+- 15 kV air discharge
Floors should consist of wood
or cement or be covered with
ceramic tiles. If the floor
consists of synthetic materials,
relative humidity must be at
least 30%.
Fast transient
electrical
disturbance/bursts
according to IEC
61000-4-4
± 1 kV
100 kHz repetition rate
± 1 kV
100 kHz repetition rate
Surge voltages
(surges) according to
IEC 61000-4-5
Not applicable
The Mobil-O-Graph
®
does not
have an AC power supply
Summary of Contents for Mobil-O-Graph PWA
Page 2: ...Revision B 17 10 2018...
Page 3: ...Gebrauchsanweisung DE Mobil O Graph PWA Mobil O Graph NG...
Page 66: ...64...
Page 67: ...Operating manual EN Mobil O Graph PWA Mobil O Graph NG...
Page 129: ...Manual de instrucciones ES Mobil O Graph PWA Mobil O Graph NG...
Page 191: ...Mode d emploi FR Mobil O Graph PWA Mobil O Graph NG...
Page 254: ...252...
Page 255: ...Istruzioni per l uso IT Mobil O Graph PWA Mobil O Graph NG...
Page 320: ...318...
Page 321: ...Gebruiksaanwijzing NL Mobil O Graph PWA Mobil O Graph NG...
Page 352: ...Productbeschrijving 350 Geschatte levensduur van de manchet 6 maanden...
Page 384: ......
Page 385: ......