Troubleshooting
105
EN
Error message
Possible cause
Measures
Err 5
bAtt
1. Power pack or battery voltage
too low
1. Replace the power packs or
batteries
2. Power packs or batteries
defective
2. The power pack or battery
voltage is correct but
bAtt
is shown
on the display during cuff inflation.
Replace the power packs
3. Battery contacts are corroded
3. Clean the battery contacts with a
cotton cloth and a little alcohol.
Err 6 +
possible
continuous alarm
until a button is
pressed
1. Build-up of air
1. Check the cuff for a build-up of
air or a kink in the tube. If there is a
kink in the cuff tube, lay the tube
out. Otherwise, send the device in
immediately.
2. Blood pressure cuff incorrectly
connected
2. Connect the cuff to the device
(see chapter 3.3.5 “Cuff
connection”)
3. Leak in the cuff or connecting
tube
3. If necessary, replace the cuff or
connecting tube.