SECTION THREE
–
COMMANDS
IDRONAUT
–
Brugherio (MB)
OCEAN SEVEN 304
Plus
CTD
12-2015
27
02 Temp 0001 254 0003
03 Cond 0002 253 0003
04 Sal 0004 255 0003
05 UNKNW 0255 255 0000
06 UNKNW 0255 255 0000
07 UNKNW 0255 255 0002
08 UNKNW 0255 255 0002
09 UNKNW 0255 255 0002
10 UNKNW 0255 255 0002
11 UNKNW 0255 255 0002
12 UNKNW 0255 255 0002
13 UNKNW 0255 255 0002
14 UNKNW 0255 255 0002
15 UNKNW 0255 255 0002
CMD:I)nitialize,M)odify,D)elete,Q)uit
At the end of the list, commands available to perform the configuration are shown:
Initialize
Completely deletes the list.
Modify
Allows the operator to enter a new sensor/parameter in the list or modify an
existing one. See below for the details about configuration.
Delete
Allows the operator to delete a configured sensor/parameter from the list.
Quit
Terminates editing and ends the set-up too.
The following parameters can be configured by means of the Modify command:
Sensor code[0..24,255=NU]
Sensor mux.[0..3,255=NU]
Sensor precision [0..6]
The below table reports the value associated to the available sensors/parameters:
Sensor Code
Sensor Description
Sensor Mux.
Sensor Precision
0
Pressure
254
2
1
Temperature
254
3
2
Conductivity
253
3
3
Conductivity @20°C
255
3
4
Salinity
255
3
5
O
2
ppm
255
2
6
O
2
%
1
1
16
Sound-V
255
4
18
Conductivity
@25°C
255
3
22
Density
255
6
24
Turbidity
0
1
29
TDS
255
3
Mux channel 255
= Derived parameter does not belong to a single sensor.
Mux channel 254/3
= Special CTD preamplifier.
The present firmware release includes the following processed/derived parameters: SALINITY,
SOUND SPEED, DENSITY, CONDUCTIVITY@ 20°C, CONDUCTIVITY@ 25°C and TOTAL