Unpack
No special handling is required for unpacking this equipment at the site. Conventional
shipping materials are used.
To remove the table from the shipping pallet, first remove the shipping brackets that
secure the table to the wooded pallet using a ½” wrench. Then roll the table down a
temporary ramp or lift off pallet with a forklift. When using a forklift, pad the forks to
protect the paint. Lift from under the metal base. When rolling the table off the pallet by
hand, use the long metal rails attached to the tabletop to guide the table after releasing the
caster brakes.
Install
Mobile tables are equipped with four locking caster brakes. Unlock brakes while positioning.
Foot controls are standard with this table. The operator should become familiar with the controls
before using.
The table operates on 110Vac (or 230Vac) or on internal battery backup. Swing the hinged cover
open to find the power cord on some models. Plug the AC power cord into a properly grounded
hospital grade outlet. The green indicator on the control box with illuminate, however, the hinged
cover will need to be opened to see the green indicator.
The outlet used should be visible and accessible to the user. The cord should be routed where it
will not be subject to damage or be a hazard. Check that the power cord ground pin is in good
condition. Check that the cord ground wire is securely connected to the ground stud inside the
unit.