SETUP
Unpack
No special handling is required for unpacking this equipment at the site. Conventional
shipping materials are used.
Roll the table down a temporary ramp (if table is mobile) or lift off pallet with a forklift. When
using a forklift, pad the forks to protect the paint. Lift from under the metal base. When rolling
the table off the pallet by hand, use the metal rails attached to the tabletop (if table is equipped
with accessory rails) to guide the table after releasing the caster brakes (if table is
mobile).
Install
Mobile tables are equipped with four locking caster brakes. Unlock brakes while positioning.
Hand controls are standard with this table with foot controls as optional. The operator should
become familiar with the controls before using.
The table operates on 110Vac (or 230Vac) or on internal battery backup. Swing the hinged cover
open to find the power cord on some models. Plug the AC power cord into a properly grounded
hospital grade outlet. The indicator on the control box with illuminate, however, depending on
table model, this indicator light may not be visible without removing a cover.
During an AC failure, the table will automatically run on battery backup without any manual
switching required. It is recommended that power be applied at all times, even when not in use, to
keep a proper charge on the batteries and achieve maximum battery life. (Batteries are being
charged during normal use).
Recycle or dispose of shipping