Operator Manual
Aspect ISO-Drive
L100-3265 Rev. A
24
Problem/Symptom
Possible Cause
Remedy
4. ISO-Centric Roll
does not operate.
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Hand Control Failure.
5. Control connection.
6. Actuator(s)
uncalibrated.
7. Actuator Failure.
1. Check electrical outlet.
2. Connect to AC Power
3. Reset Emergency Stop Pushbutton.
4. Replace Hand Control.
5. Service Control connections
according to Service Manual.
6. Reinitialize Actuator Controller with
Handheld Control. (Section 12.3)
7. Replace ISO-Roll Actuator
according to Service Manual.
5. No response to
Handset.
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Handset Control Cable
damaged.
5. Actuator(s)
uncalibrated.
6. Device Failure.
1. Check electrical outlet.
2. Connect to AC Power.
3. Reset Emergency Stop Pushbutton.
5. Reinitialize Actuator Controller with
Handset Control. (Section 12.3)
6. Replace faulty device.
6. Elevate,
Trendelenburg and
ISO-Roll slow or
unexpected
movement.
1. Running on Battery
Power.
2. Actuator(s)
uncalibrated.
1. Connect to AC Power.
2. Reinitialize Actuator Controller with
Handset Control. (Section 12.3)