Running quality control on your analyzer

The purpose of quality control is to verify the integrity of your results and to verify that your analyzer is 

functioning optimally. IDEXX recommends that you perform quality control once per month.

Note:

 Keep SediVue* QC in the refrigerator—DO NOT freeze. SediVue QC has a 12-month dating.

1.  Tap the 

SediVue Dx

 icon on the IDEXX VetLab* Station Home screen.

2.  Tap 

Quality Control

, select the QC lot you are using, and then tap 

Run QC

.

4.  Gently invert the vial 10 times to ensure the cells at the bottom of the vial are completely suspended.
5.  Using the SediVue* Pipette with a new pipette tip, aspirate 165 µL of L1 QC fluid and inject it into a 

cartridge.

6.  Press the 

Start

 button on the analyzer.

7.  Repeat the above steps for the L2 QC vial.
8.  Verify the results are in the target ranges and that the images are sharp and in focus. 

Cleaning the analyzer

To ensure the optimal performance of your SediVue Dx* Urine Sediment Analyzer, it’s recommended that you clean the internal 

components of the analyzer on a routine (monthly) or as-needed basis. 

Note: 

You’ll need alcohol prep pads and optical tissues when performing routine and preventive maintenance. Need to order more? 

Get them free at order.idexx.com (in Canada, go to idexx.ca/orderinfo).

To clean after a cartridge was overfilled or filled incorrectly

For best results, after you notice that the cartridge has been inappropriately filled, 

do not press the Start button as this will drag 

urine into the internal components of the analyzer

. Instead, follow the instructions below.

Note:

 Already pressed the Start button? Follow the instructions for routine monthly maintenance. 

1.  Open the analyzer door. Then 

with gloved hands use an optical 

tissue to absorb urine on top of the 

cartridge. 

2.  Remove and dispose of the 

cartridge.

3.  Using a new alcohol prep pad, 

clean the feed arm (dispense area), 

including the front of the move arm. 

Then dry the components with an 

optical tissue.

4.  Close the analyzer door. Once the 

SediVue Dx icon has a green, “ready” 

status, tap the patient icon on the 

IDEXX VetLab* Station Home screen 

and tap 

Start Run

. A new cartridge 

is placed in the pipetting window.

Feed arm

Move arm