Idexx SediVue Dx User Manual Download Page 2

Analyzer components

Sample preparation

Urine samples can be obtained via cystocentesis, catheterization, or free catch method and should be analyzed within  
30 minutes of collection (or stored in the refrigerator and returned to room temperature prior to analysis).

Sleeve lock

Centrifuge shield

Drip tray

Cartridge holder 

Track 

Optical stage 

Optical  
window 

If the sample is in a container:

1.  Cap the container and invert it 10 times to mix—a well-

mixed sample is essential to obtaining accurate results. 

2.  Immediately after inversion, use the SediVue* Pipette to 

aspirate 165 µL of sample from the bottom half of the 
container, ensuring there are no bubbles in the sample.

When using the pipette, follow these important guidelines:

•  Crystals settle 

immediately after 
inversion. Aspirate from 
the bottom half of the 
container.

•  Remove excess urine 

on the tip with a lint-free 
wipe.

•  Hold vertically.
•  ONLY depress plunger 

to the first stop.

•  Use only IDEXX-

supplied pipette tips.

If the sample is in a syringe:

1.  In a 1 cc–5 cc syringe 

containing at least 0.5 cc of 
urine, draw 0.5 cc of air into 
the syringe.

2.  Replace the needle with a 

SediVue* Syringe Tip. 

3.  Gently invert the syringe 10 

times to mix the sample. 

4.  Immediately after inversion, 

purge all of the air from the 
syringe and inject 165 µL of 
urine into the cartridge fill 
port.

2

3

Analyzing a sample

1.  Initiate the sample run on the IDEXX VetLab* Station.
2.  Confirm the patient information, select the 

SediVue Dx

 icon, and then tap 

Run

.

3.  Follow the sample preparation guidelines on the previous page. 

IMPORTANT—When injecting sample into the cartridge: 

•  Ensure the sample fills the cartridge window but not the overflow channel.
•  Make sure the pipette tip is firmly seated into the fill port on the cartridge.
•  Using light downward pressure, depress the plunger slowly to avoid air bubbles.
•  Only depress plunger to the first stop. 
•  Lift the pipette from the fill port before you release the plunger.

4.  Press the 

Start

 button on the front of the  analyzer. 

Waste bin

Start button

Status light

Pipetting  
window
Power button
Door access

Fan filter

Stop