Troubleshooting 

Detector status lamps 

Ready light (left) 

Power light (center)

Error light (right)

ON steady

Ready for exposure

Power on

No communication

ON blinking

Initializing or exposure  

is in process

N/A

Error

OFF

Offline

Power off

OK

Maintenance

•  Once a month, clean the air filter on the power supply unit: 

1. Turn off the equipment, and then push down on the lever at the top of the filter to detach it. 
2. Clean the air filter with a vacuum cleaner, and then reattach it to the power supply.

• Once a year, clean the computer fans and wipe down the detector’s exterior with a moistened cloth.
• Replace the battery pack when storage capacity is <60% or operable time is <100 minutes.

For more information

• Refer to the 

IDEXX ImageVue* DR50 Operator’s Guide

 (in PDF format) provided with your system. 

• Call IDEXX Diagnostic Imaging Customer Support at 1-877-433-9948.

Status lamps

メイン

スイッチ

電源状態

表示用

LED

Main switch

Power 

status LED

Air filter

Power supply unit

Intended use: The IDEXX ImageVue* DR50 system is intended for veterinary use 
only; it is not intended for human diagnostic use. For a detailed list of species and 
settings, contact IDEXX Diagnostic Imaging Customer Support: 