25
Advanced Control
In each box of Advanced Control, there is one vial containing the control fluid. The lot number can
be found on the product packaging.
Note:
Each vial contains enough fluid for 2 runs, in the event a secondary run is necessary.
Storage
Store frozen until the expiration date, or store in the refrigerator for up to 5 days.
Stability and Handling
Once opened, Advanced Control cannot be stored and reused—discard remaining fluid after
use.
PHBR Control
In each box of PHBR Control, there are six vials containing the control fluid. The lot number can be
found on the product packaging.
Storage
Store frozen until the expiration date, or store in the refrigerator for up to 7 days.
Stability and Handling
Once thawed, PHBR Control cannot be stored and reused—discard remaining fluid after use.
Quality Control CLIPs and Slides
IDEXX recommends that you perform monthly quality control testing after you have cleaned
the internal components of your analyzer. The convenient Catalyst* QC CLIP contains all of the
chemistry slides needed to perform this task. It is also recommended that you perform a quality
control for electrolytes using the Catalyst* Lyte 4 CLIP.
Run the QC CLIP and the Lyte 4 CLIP
Use the convenient QC CLIP and the Lyte 4 CLIP in conjunction with the IDEXX VetTrol Control fluid
to perform quality control on your Catalyst One analyzer. It is recommended that you wait at least 30
minutes after running any slides before running the QC CLIP.
OR
Run Individual Slides
You can use individual slides to create your own QC panel and perform a quality control test
(one slide per group). If you want to use individual slides to run quality control, we recommend a
minimum of one slide from each of the groups below.
Group 1
NH
3
Group 2
AMYL
CHOL
GLU
LAC
LIPA
TBIL
TP
TRIG
Quality Control