12
Digital Input
NOTE:
Time T must be
≥
to T
MIN
Input Operation (Generic Functions)
Sampling
T
MIN
HMS2500
Frequency
10Hz
220ms
20Hz
110ms
12"
50Hz
45ms
1" to 10"
80Hz
30ms
150Hz
15ms
Activate Auto-Calibration from a Remote Input
T
MIN
< T < 1 second = Autocalibration
T > 1 second = Autozero
Necessary conditions for this function to be enabled:
•
Calibration
enabled (
10-DIAGNOSTIC
menu).
•
Batch
on digital input 1 disabled (
5-INPUTS
menu).
•
Batch
on digital input 2 disabled (
5-INPUTS
menu).
NOTE
: This feature should not be used for applications
with large viscosity variations.
Disable Totalizer from a Remote Input
Necessary conditions for this function to be enabled:
•
Count lock
enabled (
5-INPUTS
menu).
•
Auto batch
disabled (
12-BATCH
menu).
•
Cons. mode
disabled (
12-BATCH
menu).
Activate Range Change from a Remote Input
Necessary conditions for this function to be enabled:
•
Range change
enabled (
5-INPUTS
menu).
•
Batch
on digital input 2 disabled (
5-INPUTS
menu).
•
Batch
functions assigned to digital input 2 disabled.
•
Out1
,
Out2
,
Out3
, or
Out4
batch end functions
assigned to digital output 2 and digital output 1 or 2 is
disabled (
6-OUTPUTS
menu).
Reset Totalizer from a Remote Input
Necessary conditions for this function to be enabled:
•
One of the following must be enabled:
•
T+ RESET
(
5-INPUTS
menu).
•
T- RESET
(
5-INPUTS
menu).
•
P+ RESET
(
5-INPUTS
menu).
•
P- RESET
(
5-INPUTS
menu).
(It is recommended that only the partial totalizers P+/P- be
enabled to avoid losing continuous totalizer data.)
Summary of Contents for LIQUID CONTROLS HML210
Page 43: ...43 Notes...