Features

 Three different tones selected from front panel

 Tone can be added in any status – allows test sets without monitor feature to pick up tone

 Separate Talk Battery status for increased voltage and power for test sets

 Auto-OFF of tone after three hours to conserve battery life

 Constant tone amplitude over life of battery

 Two tone amplitude – normal and half

 No power draw in continuity status with leads open

 RJ-11 jack allows use of modular cable or alligator clips

 Low battery LED indicates when to change battery

 9 Volt battery included

 F-Connector port for toning and tracing of coax cable

  

Warning!

Do not attach to live AC circuits. This could cause an extreme shock hazard and damage to the Tone Generator.  

When connecting to unknown circuits, always start in the OFF Status. If either or both status LEDS light,  

there is voltage present. The CONT and TALK BATT status modes should NOT be used with voltage present.

Instructions for Use

To send a tone for tracing

  Caution: Do not connect to an active AC circuit exceeding 24V in this mode

1. Move slide switch to “OFF” Status position

2. Connect black lead to Tip/positive/white striped conductor.

3. Connect red lead to Ring/negative/solid colored conductor

4. PRESS and RELEASE Tone Button to generate FULL Strength tone

5. PRESS and RELEASE Tone Button again to generate HALF-Strength tone

6. PRESS and HOLD Tone Button to cycle through three available tones: warble, pulse and steady

7.  Use IDEAL Amplifier Probe (62-164) to receive tone. Reception of tone will be strongest on the  

subject wires. In cases of ready access to bare conductors, a handset may be used to receive tone.

8. To turn tone generator off, PRESS and RELEASE Tone Button until LEDs do not glow.

Application Tips

  For the strongest possible tone strength, it is recommended to connect one lead to a conductor and the other 

lead to a ground. If no ground is available, dangle the second lead as close to earth as possible. 

(See Fig. 1)

 

 With Tone on, the CONT status mode can be used to verify the correct pair at the remote end.

  To confirm correct pair at the remote end, short out the pair with the strongest tone received by touching the leads 

together. The tone will “cancel” out or stop, and the REV LED will light on the tone generator indicating continuity.

To check continuity

  Caution: Do not connect to live AC circuits

1. Move slide switch to “CONT” position

2. Connect leads to subject pair

3. Red LED indicates continuity, no LED indicates open or line resistance over 10,000 ohms.

4. Move slide to “OFF” Status to avoid draining battery of leads touch during storage

Providing Talk Power  

Caution: Do not connect to live AC circuits

1. Move slide switch to “TALK BATT” position

2. Connect tone generator in series with one test set at each end of a dead pair, creating a closed loop.

3. Place both test sets “off hook” or TALK position to establish communication. 

(See Fig. 2)

Analog Telephone Line Testing 

(Central Office Battery must be present to perform tests)

   

Caution: Do not connect to to live AC circuits

Telephone Line Polarity

1. Move slide switch to “OFF Status” position

2. Connect black lead to Tip/positive/white striped conductor.

3. Connect red lead to Ring/negative/solid colored conductor.

4. If NRM LED glows, polarity is correct. If REV LED glows, polarity is reversed.

Telephone Line Status

1. Move slide switch to “OFF Status” position

2. Connect black lead to Tip/positive/white striped conductor.

3. Connect red lead to Ring/negative/solid colored conductor.

  a. If NRM or REV LED is bright, the line is not in use

  b. If NRM or REV LED is dim, the circuit is in use

  c. If both LED’s are flashing, ringing current is present

  d. If both LED’s glow, AC Voltage is present

Modular Testing

All above tests are available through the modular plug for Line 1 only – red and green wires / blue pair

Coax Testing

1.  To test unterminated coax, connect red to outer shield and black to center conductor OR  

red to outer shield and black to ground

2.  To test terminated coax, connect red to connector housing and black to center pin OR  

red to connector and black to ground

Battery Replacement

1. Unscrew battery compartment 

(See Fig. 3)

2. Remove and replace battery

3. Replace screw on battery compartment

T

O

N

E

HI

Dua

l

(HO

LD)

Sin

gle

NRM

REV

LO

TALK 

 

  OF

F

BA

TT  CONT  

S

TA

TU

S

 62-1

64

NRM

REV

HI

LO

OF
F

TO

NE

GENERA

TO

R

 62-
160

T

O

N

E

NRM

REV

HI

Dual

(HOLD)

Single

LO

TA

LK 
 

  OF
F

B

A

TT 
 CONT  ST

A

T

US

TO

NE

GENERA

TO

R

CA
T NO

.

 

62-16

0

Made 

in

 Chin

a

Tone Generator 

Instruction Sheet

Specifications

Power requirements

 

One 9 Volt battery

Battery Life 

70 hours typical

Tone Frequencies

 

Dual/1116 and 919 Hz,  

Single/919 Hz

Tone Power into 600 ohms

 

HI=7dBm, LO=1dBm

Voltage Protection 

OFF Status – DC=58 volts  

continuous, AC=300Vpeak,  

2 seconds Continuity and  

Talk Battery into 600 ohms –  

DC=52 volts 

Talk Battery into 600 ohms

 

6.0 VDC

Dimensions

 

2.7 x 2.6 x 1.1 inches  

(67 x 66 x 28 mm)

Weight

 

4 oz with battery (113 grams)

Fig. 1

Fig. 2

Fig. 3

TONE BUTTON

DUAL  

FUNCTION LED’s

MODE SWITCh

TELEPhONE LINE  

STATUS LED’s

BATTERY COMPARTMENT

Características

 Se pueden escoger tres tonos diferentes en el tablero delantero

  Se puede añadir tono en cualquier estado – permite que los aparatos de prueba sin característica  

de supervisión capten el tono 

 Estado de pila de conversación separado para mayor voltaje y corriente para los aparatos de prueba

 Desconexión automática del tono después de tres horas para conservar la duración de la pila

 Amplitud de tonos constante a lo largo de la duración de la pila

 Dos amplitudes de tonos – normal y medio

 No se consume corriente en estado de continuidad con los cables abiertos

 Jack RJ-11 que permite el uso de un cable modular o cables de pinza de cocodrilo 

 LED de pila poco cargada que indica cuándo hay que cambiar la pila

 Se incluye una pila de 9 voltios

 Zócalo para conector F para envío de tonos y rastreo de cables coaxiales

  

¡Advertencia!

No lo conecte a circuitos activos de CA. Esto puede causar un peligro extremo de electrocución  

y daños en el generador de tonos Al conectarlo a circuitos desconocidos, empiece siempre en  

el estado APAGADO. Si se enciende uno o dos LED, no hay voltaje presente. NO se deben usar  

las modalidades de estado CONT y TALK BATT con voltaje presente.

Instrucciones de empleo

Para enviar un tono para rastrear   

Precaución: No lo conecte a un circuito de CA activo que exceda 24V en esta modalidad

1. Mueva el interruptor deslizante al estado “APAGADO” 

2. Conecte el cable negro al conductor de punta/positivo/con franjas blancas.

3. Conecte el cable rojo al conductor anillo/negativo/de un solo color

4. PULSE y SUELTE el botón del tono para generar el tono de intensidad MÁXIMA

5. PULSE y SUELTE nuevamente el botón del tono para generar el tono de intensidad MEDIA

6.  PULSE y MANTENGA PULSADO el botón del tono para ciclar los tres tonos disponibles:  

ululado, impulsos y constante

7.  Use la sonda amplificadora de IDEAL (62-164) para recibir el tono. La recepción de tono será más intensa en los cables 

tratados. En casos de acceso listo a conductores descubiertos, se puede usar un aparato de pruebas para recibir el tono.

8. Para apagar el generador de tonos, PULSE y SUELTE el botón del tono hasta que se apaguen los LED.

Recomendaciones de aplicación

  Para la máxima intensidad posible del tono, se recomienda conectar un cable a un conductor y el otro cable a tierra.  

Si no hay una conexión a tierra, deje colgar el segundo cable lo más cerca posible de tierra. 

(Vea la Fig. 1)

  Con el tono conectado, se puede usar la modalidad de estado CONT para verificar el par correcto en el extremo remoto.

  Para confirmar el par correcto en el extremo correcto, cortocircuite el par con el tono más fuerte recibido  

poniendo en contacto los cables. El tono se “cancelará” o cesará, y el LED REV del generador de tonos se  

encenderá indicando continuidad.

Para comprobar la continuidad  

Precaución: No lo conecte con circuitos activos de CA.

Generador de tones 

Hoja de instrucciones

Especificaciones

Requisitos de alimentación

 

Una pila de 9 voltios

Duración de la pila 

70 horas típico

Frecuencias de tono

 

Doble/1116 y 919 Hz,  

sencillo/919 Hz

Corriente de tono  

en 600 ohmios

 

ALTO=7dBm, BAJO=1dBm

Protección del voltaje 

Estado APAGADO – CC=58 voltios 

continuos, CA=300V máximo,  

2 segundos de continuidad y  

pila de conversación en600 

ohmios – CC=52 voltios 

Pila de conversación  

en 600 ohmios

 

6,0 VCC

Dimensiones

 

2,7 x 2,6 x 1,1 pulg  

(67 x 66 x 28 mm)

Peso

 

4 onzas con pila (113 gramos)

 62-1

64

NRM

REV

HI

LO

OF
F

TO

NE

GENERA

TO

R

 

Fig. 1

F-CONNECTOR PORT

T

O

N

E

HI

Dua

l

(HO

LD)

Sin

gle

NRM

REV

LO

TALK 

 

  OF

F

BA

TT  CONT  

S

TA

TU

S

BOTÓN DE TONO 

LED INDICADORES  

DE DOBLE FUNCIÓN 

INTERRUPTOR 

DE MODO

LED INDICADORES  

DE ESTADO DE LA  

LÍNEA TELEFÓNICA 

COMPARTIMIENTO  

DE LA PILA 

ZÓCALO PARA CONECTOR F 

Reviews: