joints, mesh, clips, orthopedic hardware).

•  DO NOT

 use if patients are taking medications such as anti-aggregation medications (NSAIDS), Isotreti-

noin (Roaccutane), or rosuvastatin (Crestor).

•  DO NOT

 use the device at all on, near, or around implanted electrical medical devices (examples: pace-

makers, insulin pumps).

•  DO NOT

 use treatment gels, oils or solutions that have alcohol or a highly flammable substance as the 

base carrier. Ignition may occur. Avoid letting these materials pool or collect in depressions or body 

cavities, such as the navel. If alcohol is used for cleaning purposes, allow thorough drying before device 

use.

•  WARNING: 

Device contains high-voltage components. Changes or modifications to the equipment 

not expressly approved by Ideal Curves or described in this manual could void the user’s authority to 

operate the equipment. Unauthorized alterations may result in electric shock or damage to patient 

tissues.

•  WARNING: 

To avoid the risk of electric shock, this equipment must only be connected to a wall outlet 

or surge protector with a third ground pin socket. Failure of the equipment may resulted in an unin-

tended increase in output power.

•  WARNING:

 Use of the Ultimate Contour adjacent to or stacked with other equipment should be 

avoided because it could result in improper operation. If such use is necessary, the Ultimate Contour 

and the other equipment should be observed to verify that they are operating normally.

•  WARNING: 

Use of accessories and cables other than those specified for the Ultimate Contour or 

provided by Ideal Curves could result in increased electromagnetic emissions or decreased electromag-

netic immunity of the Ultimate Contour and result in improper operation.

•  WARNING:

 Portable RF communications equipment (including peripherals such as antenna cables 

and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Ultimate  

Contour, including cables specified by Ideal Curves. Otherwise, degradation of the performance of this 

equipment could result.

•  CAUTION: 

Use of ultrasound on body parts other than the waist, or on tissue that has a low adipose or 

fatty content may result in rapid heating, tissue damage, and/or the formation of fistulas.

•  CAUTION: 

DO NOT store liquids on the unit.

•  CAUTION: 

Use of controls or adjustments or performance of procedures other than those specified 

herein may result in hazardous exposure to radio frequency or ultrasound energy.

•  CAUTION: 

Always stop the treatment before removing the handpiece from contacting the patient. 

Removing the handpiece while the energy is emitting may result in electric shock or damage to the 

tissue.

•  CAUTION: 

Federal law restricts this device to sale by or on the order of a physician.

•  Device is only to be used in a professional medical office environment, where environmental controls 

are in place. Specifically, the device should not be used in a dry or arid environment where the relative 

humidity in the room is < 35%. See Section 9.1.

•  When using the ultrasound handpiece, gas bubbles may form at the surface of the handpiece. Ensure 

that treatment oils are used that are suitable for ultrasound use to maintain contact between the 

handpiece and the skin (see page 9).

•  Locate the device on a table, counter, or cart when in use. DO NOT place the device on the floor when 

using.

•  Ensure all other radio frequency emitting devices (example: cell phones, computers and wireless 

computer devices, routers, wireless phones) or devices that may generate a strong electromagnetic 

field (example: MRI, electrosurge) are kept more than 3 meters (10 feet) away from this device, the 

treatment table, the operator, or the patient.

•  Avoid use of the device if liquids are spilled on the unit. Wipe up spills immediately and ensure the unit 

Summary of Contents for Ultimate Contour

Page 1: ......

Page 2: ...p 9 4 3 Registration 9 4 4 Handpiece Attachment 9 5 DEVICE OPERATION 10 5 1 Patient Preparation 10 5 2 Screens and Settings 10 5 3 Radio Frequency Treatment 13 5 4 Ultrasound Treatment 14 5 5 Concludi...

Page 3: ...y Failure to read these instructions may result in damage to the device or risk of injury to you or to the patient It is advisable to keep this document with the device and refer to it often when ques...

Page 4: ...s with a Body Mass Index BMI under 25 DO NOT use radio therapy on or near the face or neck DO NOT use on pregnant or nursing women on those who have given birth in the last 12 months or on those plann...

Page 5: ...ls such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the Ultimate Contour including cables specified by Ideal Curves Otherwise degradation of th...

Page 6: ...onnected to an oxygen supply DO NOT use in the presence of flammable gases All treatments performed with this system must be subjected to the same clinical assessment as with traditional techniques Th...

Page 7: ...e waist circumference of patients using the ultrasound portion This study evaluated a total of 42 individuals who were treated for 30 minutes at an intensity level of 4 during each treatment session T...

Page 8: ...hock ONLY Federal law restricts this device to sale by or on the order of a physician CAUTION Make sure the unit is positioned so you can get easy access to the back panel of the unit in case you need...

Page 9: ...e easily disconnected if needed If needed clean the unit and handpieces as described in Section 6 4 3 Registration Before beginning use of the Ultimate Contour it is highly recommended that you take a...

Page 10: ...istent with the diameter of the handpiece This will help ensure that all of the treatment area is addressed and to help avoid overlapping treatment spots applying excessive energy to a section of tiss...

Page 11: ...rently selected The relative power level of the energy The handpiece that is connected to the unit Details about specific differences will be discussed in Sections 5 3 To change the treatment time or...

Page 12: ...stop the treatment early and then press the Standby button to return to the Standby screen the time will reset to the full amount of time you had selected CAUTION Do not lift up the handpiece during...

Page 13: ...rth sweeping motion A good analogy to think of is mowing a lawn consistent sweeping patterns that just barely overlap slowly working from one side of the area to the other Apply slight pressure into t...

Page 14: ...and the unit has stopped delivering energy lift the handpiece away from the patient tissue Set the handpiece aside Use clean towels to wipe away excess treatment oil from the skin Assess the treated a...

Page 15: ...be treated with a disinfection solution such as a 1 3 glutaraldehyde solution or 0 2 peracetic acid solution The handpiece should be visibly wet and remain so for at least 5 minutes You may need to re...

Page 16: ...th its accessories excluding the handpieces collectively defined as Product is warranted against defective materials and workmanship for a period of one 1 years from the date of purchase If the Produc...

Page 17: ...you are having trouble with the Ultimate Contour system have questions regarding its use or need to order replacement parts feel free to contact a customer service representative at Ideal Curves 1 888...

Page 18: ...ins of a burning sensation Some warmth is normal as this indicates the procedure is working If the heat is intolerable use a lower power setting Also ensure an ample amount of treatment oil is applied...

Page 19: ...0 50 W cm2 1 00 W cm2 Level 3 0 75 W cm2 1 50 W cm2 Level 4 1 05 W cm2 2 00 W cm2 Ultrasound Focal Distance 6mm Ultrasound Focal Distance 6mm Ultrasound Focal Volume 2 12cm3 Ultrasound Entry Power 30...

Page 20: ...to take mitigation measures such as relocating or re orienting the equipment Harmonic emissions IEC 61000 3 2 Not available Voltage Fluctuations Flicker emissions Not available The Ultimate Contour i...

Page 21: ...in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey...

Page 22: ...ration distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 d 3 5 V1 P d 3 5 E1 P GHz d 7 E1 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10...

Page 23: ...23 THIS PAGE INTENTIONALLY LEFT BLANK...

Page 24: ...Manufactured for Ideal Curves LLC by CAO Group 4628 West Skyhawk Dr West Jordan UT 84084 www ultimate contour com Rev 4 0 Mar 2021 1 888 861 7211...

Reviews: