CAUTION: 

Make sure to select the minimum power level that will achieve the intended results.  

The maximum power setting presents increased risk of thermal damage relative to the lower 

settings.

Position the handpiece onto the patient at the first point of exposure. Make sure the skin is 

                          still coated with the treatment oil.

CAUTION: 

 Ensure that the handpiece is in direct contact with the skin prior to starting the 

treatment. Air gaps between the handpiece and the tissue may result in an electric shock.

Remind the patient that they may experience a warming sensation during the treatment, 

and advise them to indicate if the handpiece feels unbearably hot during the treatment. With 

everything prepared, press the Start button on the screen to begin the treatment.

During treatment, the Stop button will be present on the screen. The Time indicator will start counting 

down. The device will emit a beep every 10 seconds during countdown. If you need to interrupt treatment 

for any reason, or need to lift the handpiece to treat a different area, press the Stop button. Reposition 

the handpiece as needed, then once the handpiece is in contact with skin again press the Start button 

to resume the treatment. The unit will continue counting down time where you left off. If you stop the 

treatment early, and then press the Standby button to return to the Standby screen, the time will reset to 

the full amount of time you had selected.

CAUTION: 

 Do not lift up the handpiece during treatment. Stop the procedure first before lifting up the 

handpiece. Do not touch the patient with your hands while the handpiece is active.

When 15 seconds remains in the treatment time, the unit will start to beep once a second indicating the 

treatment is almost over. Once the timer reaches 0, the unit will automatically deactivate the handpiece 

and the system will show the Ready screen. The time will reset to the total that you had originally selected. 

Once the treatment is ended, you can lift up the handpiece.

 

Media Screen

From this screen you can access a number of tutorial videos that 

demonstrate use of the Ultimate Contour and address some fre-

quently asked questions. You can also find an electronic version 

of this Owner’s Manual for the Ultimate Contour. Press on any of 

the buttons to launch the video, or view the manual. When done 

viewing the video or manual, press the Back button at upper left 

to return to the Media Screen.

Utilities Screen

Here you can adjust the volume of the sounds made by the 

Ultimate Contour unit. Simply press the Volume button to cycle 

through the available volume levels. This screen also provides 

contact information for Ideal Curves in case of questions, as well 

as the software revision information for the Ultimate Contour.

Summary of Contents for Ultimate Contour

Page 1: ......

Page 2: ...p 9 4 3 Registration 9 4 4 Handpiece Attachment 9 5 DEVICE OPERATION 10 5 1 Patient Preparation 10 5 2 Screens and Settings 10 5 3 Radio Frequency Treatment 13 5 4 Ultrasound Treatment 14 5 5 Concludi...

Page 3: ...y Failure to read these instructions may result in damage to the device or risk of injury to you or to the patient It is advisable to keep this document with the device and refer to it often when ques...

Page 4: ...s with a Body Mass Index BMI under 25 DO NOT use radio therapy on or near the face or neck DO NOT use on pregnant or nursing women on those who have given birth in the last 12 months or on those plann...

Page 5: ...ls such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the Ultimate Contour including cables specified by Ideal Curves Otherwise degradation of th...

Page 6: ...onnected to an oxygen supply DO NOT use in the presence of flammable gases All treatments performed with this system must be subjected to the same clinical assessment as with traditional techniques Th...

Page 7: ...e waist circumference of patients using the ultrasound portion This study evaluated a total of 42 individuals who were treated for 30 minutes at an intensity level of 4 during each treatment session T...

Page 8: ...hock ONLY Federal law restricts this device to sale by or on the order of a physician CAUTION Make sure the unit is positioned so you can get easy access to the back panel of the unit in case you need...

Page 9: ...e easily disconnected if needed If needed clean the unit and handpieces as described in Section 6 4 3 Registration Before beginning use of the Ultimate Contour it is highly recommended that you take a...

Page 10: ...istent with the diameter of the handpiece This will help ensure that all of the treatment area is addressed and to help avoid overlapping treatment spots applying excessive energy to a section of tiss...

Page 11: ...rently selected The relative power level of the energy The handpiece that is connected to the unit Details about specific differences will be discussed in Sections 5 3 To change the treatment time or...

Page 12: ...stop the treatment early and then press the Standby button to return to the Standby screen the time will reset to the full amount of time you had selected CAUTION Do not lift up the handpiece during...

Page 13: ...rth sweeping motion A good analogy to think of is mowing a lawn consistent sweeping patterns that just barely overlap slowly working from one side of the area to the other Apply slight pressure into t...

Page 14: ...and the unit has stopped delivering energy lift the handpiece away from the patient tissue Set the handpiece aside Use clean towels to wipe away excess treatment oil from the skin Assess the treated a...

Page 15: ...be treated with a disinfection solution such as a 1 3 glutaraldehyde solution or 0 2 peracetic acid solution The handpiece should be visibly wet and remain so for at least 5 minutes You may need to re...

Page 16: ...th its accessories excluding the handpieces collectively defined as Product is warranted against defective materials and workmanship for a period of one 1 years from the date of purchase If the Produc...

Page 17: ...you are having trouble with the Ultimate Contour system have questions regarding its use or need to order replacement parts feel free to contact a customer service representative at Ideal Curves 1 888...

Page 18: ...ins of a burning sensation Some warmth is normal as this indicates the procedure is working If the heat is intolerable use a lower power setting Also ensure an ample amount of treatment oil is applied...

Page 19: ...0 50 W cm2 1 00 W cm2 Level 3 0 75 W cm2 1 50 W cm2 Level 4 1 05 W cm2 2 00 W cm2 Ultrasound Focal Distance 6mm Ultrasound Focal Distance 6mm Ultrasound Focal Volume 2 12cm3 Ultrasound Entry Power 30...

Page 20: ...to take mitigation measures such as relocating or re orienting the equipment Harmonic emissions IEC 61000 3 2 Not available Voltage Fluctuations Flicker emissions Not available The Ultimate Contour i...

Page 21: ...in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey...

Page 22: ...ration distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 d 3 5 V1 P d 3 5 E1 P GHz d 7 E1 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10...

Page 23: ...23 THIS PAGE INTENTIONALLY LEFT BLANK...