9

outlet. Use of the Ultimate Contour in abnormally hot or cold temperatures should be avoided. Refer to 

Section 9.1 for the preferred operating temperature. If the unit has been stored or transported in extreme 

conditions, allow time for the unit to acclimate to room temperature before turning it on and using it.

Ensure that the treatment table the patient will be on has no exposed metal surfaces that the patient may 

come in contact with during the treatment.

 

4.2 Device Unpacking and Setup

When removing the Ultimate Contour unit and accessories, take care to avoid scratching or damaging any 

of the components. 

•  Remove the Ultimate Contour unit from its packaging, and stand the unit upright on a sturdy table or 

countertop surface. 

•  Remove the handpieces from the packaging and place these in proximity to the unit. Take care not to 

twist the transducer cable or bend it in a tight loop.

•  Remove the power cord from the packaging. Plug the power cord into the back of the unit. Make sure 

the power switch on the back of the unit is turned OFF (press on the “O” symbol). Locate a suitable 

power outlet as described in Section 4.1 above and plug in the power cord. Make sure the unit is posi-

tioned so that the power cord may be easily disconnected if needed.

•  If needed, clean the unit and handpieces as described in Section 6.

4.3 Registration

Before beginning use of the Ultimate Contour, it is highly recommended that you take a moment to 

register at

 

www.ultimate-contour.com. Registration facilitates customer service, provides additional infor-

mation regarding the use of the Ultimate Contour, managing patient expectations, and other useful infor-

mation.

4.4 Handpiece Attachment

Prior to use on a patient, ensure the desired handpiece has been cleaned consistent with the instructions 

in Section 6.1. When attaching the handpiece connector to the transducer cable, pay careful attention to 

the alignment of the connectors. The connectors are designed to only attach in a particular “clock position”. 

If you find the handpiece connector does not attach, DO NOT force it. Slightly rotate the connector and try 

again until the indexing feature aligns and the connector clicks into place. To detach, squeeze the silver tab 

on the connector while gently pulling the connector out.

Make sure you attach the handpiece you intend to use. The Ultimate Contour is designed to detect which 

handpiece is attached (ultrasound or RF) and limit access to controls accordingly. This system is designed 

for Bipolar RF emissions. A neutral electrode is not necessary.

From 

Handpiece

Alignment and 

locking feature

Handpiece connector attaches 

here, with the locking feature up

Summary of Contents for 005-00036

Page 1: ......

Page 2: ...up 9 4 3 Registration 9 4 4 Handpiece Attachment 9 5 DEVICE OPERATION 10 5 1 Patient Preparation 10 5 2 Screens and Settings 10 5 3 Radio Frequency Treatment 13 5 4 Ultrasound Treatment 14 5 5 Concluding the Treatment 14 6 CLEANING AND MAINTENANCE 15 6 1 Handpiece Cleaning and Inspection 15 6 2 Unit Cleaning 15 6 3 Fuse Replacement 16 7 LIMITED WARRANTY 16 7 1 Limited Warranty Statement 16 7 2 Ser...

Page 3: ...ly Failure to read these instructions may result in damage to the device or risk of injury to you or to the patient It is advisable to keep this document with the device and refer to it often when questions arise regarding its use and care If at any point you have any questions about the use or application of the Ultimate Contour feel free to contact a customer service representative at 1 888 861 ...

Page 4: ...ns with a Body Mass Index BMI under 25 DO NOT use radio therapy on or near the face or neck DO NOT use on pregnant or nursing women on those who have given birth in the last 12 months or on those planning to become pregnant DO NOT use near plastic or metal parts within the body example implants stents ports Do not allow the transducers to contact hard surfaces when activated DO NOT use on patients...

Page 5: ...als such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the Ultimate Contour including cables specified by Ideal Curves Otherwise degradation of the performance of this equipment could result CAUTION Use of ultrasound on body parts other than the waist or on tissue that has a low adipose or fatty content may result in rapid heating tissue damag...

Page 6: ...connected to an oxygen supply DO NOT use in the presence of flammable gases All treatments performed with this system must be subjected to the same clinical assessment as with traditional techniques The patient risk and medical history must be considered and understood before treatment Caution must be exercised regarding general medical conditions not already mentioned which might contraindicate a...

Page 7: ...he waist circumference of patients using the ultrasound portion This study evaluated a total of 42 individuals who were treated for 30 minutes at an intensity level of 4 during each treatment session There were a total of 3 treatment sessions with 5 7 days interval between sessions The results show that on average a total waist reduction for the entire regimen was 2 0 inches The study included a 4...

Page 8: ...Shock ONLY Federal law restricts this device to sale by or on the order of a physician CAUTION Make sure the unit is positioned so you can get easy access to the back panel of the unit in case you need to quickly turn the unit off or unplug it 4 INSTALLATION AND SET UP 4 1 Office Conditions The Ultimate Contour should be used in a controlled medical office environment The Ultimate Contour should b...

Page 9: ...be easily disconnected if needed If needed clean the unit and handpieces as described in Section 6 4 3 Registration Before beginning use of the Ultimate Contour it is highly recommended that you take a moment to register at www ultimate contour com Registration facilitates customer service provides additional infor mation regarding the use of the Ultimate Contour managing patient expectations and ...

Page 10: ...sistent with the diameter of the handpiece This will help ensure that all of the treatment area is addressed and to help avoid overlapping treatment spots applying excessive energy to a section of tissue which could cause thermal damage Make sure to use a marker that will not get wiped away by the treatment oils but will easily clean off after the treatment Use a tape measure and or skinfold calip...

Page 11: ...rrently selected The relative power level of the energy The handpiece that is connected to the unit Details about specific differences will be discussed in Sections 5 3 To change the treatment time or the power press the Up arrow to increase the value or the Down arrow to decrease the value Minutes are selected in quantities of 10 minute incre ments from 10 to 40 minutes Intensity is selected at l...

Page 12: ...u stop the treatment early and then press the Standby button to return to the Standby screen the time will reset to the full amount of time you had selected CAUTION Do not lift up the handpiece during treatment Stop the procedure first before lifting up the handpiece Do not touch the patient with your hands while the handpiece is active When 15 seconds remains in the treatment time the unit will s...

Page 13: ...th sweeping motion A good analogy to think of is mowing a lawn consistent sweeping patterns that just barely overlap slowly working from one side of the area to the other Apply slight pressure into the skin so the handpiece has a sort of kneading effect on the tissue Most effective results is achieved when folds of skin press up in between the nodes on the handpiece surface Once the entire area ha...

Page 14: ... and the unit has stopped delivering energy lift the handpiece away from the patient tissue Set the handpiece aside Use clean towels to wipe away excess treatment oil from the skin Assess the treated area There may be some redness or slight swelling of the treatment area Capture post treatment photos of the treated area using the same camera settings as were used for the pre treatment assess ment ...

Page 15: ... be treated with a disinfection solution such as a 1 3 glutaraldehyde solution or 0 2 peracetic acid solution The handpiece should be visibly wet and remain so for at least 5 minutes You may need to reapply the disinfection solution to achieve this After the five minutes use a new dry towel or cloth to pat dry the handpiece Ensure the handpiece is fully dry before next use After each use look over...

Page 16: ...ith its accessories excluding the handpieces collectively defined as Product is warranted against defective materials and workmanship for a period of one 1 years from the date of purchase If the Product is leased from Ideal Curves or from any of its authorized distributors consult your lease agreement for the terms and conditions of Product repair and or replacement Repair and or replacement shall...

Page 17: ... you are having trouble with the Ultimate Contour system have questions regarding its use or need to order replacement parts feel free to contact a customer service representative at Ideal Curves 1 888 861 7211 or at the distributor from whom you obtained the system You may find it quick and helpful to review the Troubleshooting section of this manual Section 8 THIS SECTION INTENTIONALLY LEFT BLAN...

Page 18: ...ains of a burning sensation Some warmth is normal as this indicates the procedure is working If the heat is intolerable use a lower power setting Also ensure an ample amount of treatment oil is applied to the treatment area Tissues with less adipose tissue may get hotter relative to highly fatty tissue Adjust the power intensity accordingly Treated tissue is red or appears swollen Slight redness i...

Page 19: ... 0 50 W cm2 1 00 W cm2 Level 3 0 75 W cm2 1 50 W cm2 Level 4 1 05 W cm2 2 00 W cm2 Ultrasound Focal Distance 6mm Ultrasound Focal Distance 6mm Ultrasound Focal Volume 2 12cm3 Ultrasound Entry Power 30 2W Ultrasound Spatial Peak to Temporal Average Intensity 2 01W cm2 Ultrasound Pre Focal Temporal Average Intensity 1 32W cm2 Ultrasound Ratio Peak to Temporal Average Intensity 2 0 Ultrasound Peak Ra...

Page 20: ...d to take mitigation measures such as relocating or re orienting the equipment Harmonic emissions IEC 61000 3 2 Not available Voltage Fluctuations Flicker emissions Not available The Ultimate Contour is suitable for use in all establishments other than domestic and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Guidance an...

Page 21: ... in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher freq...

Page 22: ...aration distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 d 3 5 V1 P d 3 5 E1 P GHz d 7 E1 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 69 3 69 7 39 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicabl...

Page 23: ...23 THIS PAGE INTENTIONALLY LEFT BLANK ...

Page 24: ...Manufactured for Ideal Curves LLC by CAO Group 4628 West Skyhawk Dr West Jordan UT 84084 www ultimate contour com Rev 4 0 Mar 2021 1 888 861 7211 ...

Reviews: