10 - ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacturer’s declaration – electromagnetic emissions
The Ultimate Contour is intended for use in the electromagnetic environment specified below.
The customer or the user of the Ultimate Contour should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions
CISPR 11
Group 1
The Ultimate Contour uses RF energy to achieve its intended function. This energy exceeds
the limits identified in this standard. This product is designed to comply with this standard
only when switched on and in an idle state (not intentionally generating RF). RF emissions
are allowed to exceed limits only in the stated RF frequency range for this device.
RF emissions
CISPR 11
Class A
NOTE: The emissions characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment
(for which CISPR 11 class B is normally required) this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
Harmonic
emissions
IEC 61000-3-2
Not available
Voltage
Fluctuations/
Flicker emissions
Not available
The Ultimate Contour is suitable for use in all establishments other than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic purposes
Guidance and manufacturer’s declaration – electromagnetic emissions
The Ultimate Contour is intended for use in the electromagnetic environment specified below.
The customer or the user of the Ultimate Contour should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30 %.
Electrical fast transient/
burst
±2 kV common
mode
±2 kV for power supply
Mains power quality should be that of a typical commercial
or hospital environment.
IEC 61000-4-4
±1 kV for input/
output lines
Lines
Not Applicable
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential mode
±2 kV common mode