19
9 - SPECIFICATIONS
9.1 Unit Technical Specifications
Ultimate Contour (Model: 005-00036)
Weight
• Ultimate Contour 5.0 lbs
Dimensions
• Ultimate Contour 13.75” x 14.5” x 14”
Power Input Requirements 100-240 VAC @ 50-60Hz, 1.8A
Audible and Visual Notification Yes
Medical Electrical Equipment Classification Class I
Power Supply Cord Length (total) 8 ft.
Energy Frequency
• Radio 1.0±0.05MHz
• Ultrasound 37±2kHz
Maximum Handpiece Output Voltage
• Radio Frequency (All sizes) 250 Volts
• Ultrasound 470 Volts
Maximum Output Power
Radio Frequency
• Level 1 - 85W
• Level 2 - 100W
• Level 3 - 130W
• Level 4 - 160W
Ultrasound Output Intensity
Handpiece Surface (Entry Intensity) At the Focal Point
• Level 1 - 0.25 W/cm
2
0.50 W/cm
2
• Level 2 - 0.50 W/cm
2
1.00 W/cm
2
• Level 3 - 0.75 W/cm
2
1.50 W/cm
2
• Level 4 - 1.05 W/cm
2
2.00 W/cm
2
Ultrasound Focal Distance 6mm
Ultrasound Focal Distance 6mm
Ultrasound Focal Volume 2.12cm
3
Ultrasound Entry Power 30.2W
Ultrasound Spatial Peak to
Temporal-Average Intensity 2.01W/cm
2
Ultrasound Pre-Focal
Temporal-Average Intensity 1.32W/cm
2
Ultrasound Ratio Peak to
Temporal-Average Intensity 2.0
Ultrasound Peak
Rarefaction Pressure 117µPa
Ultrasound -6dB Beam Area 7.07 cm
2
Handpiece Voltage Rating
• Radio Frequency (All sizes) 300 Volts
• Ultrasound 1000 Volts
Operating Temperature 10 – 30°C (50 – 86°F)
Storage Temperature -20 – 45°C (4 – 113°F)
Operating Humidity 35 – 60% Relative Humidity
Storage Humidity 5 – 95% Relative Humidity
Warranty
•
Handpiece
•
Unit and all other parts of the system
1 year
1 year
9.2 Accessories and Replacements
Handpiece - Large RF
005-00040
Handpiece - Ultrasound
005-00037
Fuses (Pack of 2)
005-00041
Summary of Contents for 005-00036