7 - Led SET 

no function for temeprature calibrator 
 

8 - Led AT/ST 

no function for temeprature calibrator 
 

9 – Display PV 

It normally indicates the temperature of the heating block. 
During programming it indicates SP1 (Set Point 1)  
 

10 - Display SV 

It normally indicates the active Set  value of the heating block 
 

3.6   Switching on of the calibrator 

 

After switching on the mains power switch the controller starts with initialisation. 
The upper display (PV) indicates tESt. 
The lower display indicates the actual version of the controller e.a. rL 1.2 

After 5 sec. the initialisation is complete and the standard mode is selected automatically.  
The upper display (PV) indicates the actual heating block temperature 
The lower display (SV) indicates the actual set value of the heating block 
The installed electrical heating cartridge heats the brass block up from ambient temperature to the adjusted set 
point value.  

 

3.7   Indication of the block temperature and set temperature 

The red, 4-digit, 7-segment LED (PV) shows continuously the temperature of the heating block. If the,  in the 
green , 4-digit, 7-segment LED (SV), adjusted set temperature in the heating block is reached, the heating 
energy radiated from the block is dispensed by short switching-on pulses of the heating cartridge, so that the 
temperature inside of the block is kept constant. 
 

3.8   Adjustment of the block temperature 

The switch-on time of the heating is indicated by the red LED OUT1. 
While during the heating-up period a permanent light shows the supply of heating energy. The flashing of the 
LED shows that the temperature of the heating block is close to the adjusted set temperature, and therefore the 
heating energy is now only supplied in short intervalls. 
The control is carried out as a PID control.  
 

 

 

 
 
In order to obtain good stability of the temperature, the cycle time “tcr1” has a low value with a very frequent 
intervention of the control output. 

 

 

Summary of Contents for SIKA TP 18 200 E

Page 1: ...erating position 4 3 3 Adapter sleeves 4 3 4 Control of temperature sensors 4 3 5 Front panel description 4 3 6 Switching on of the calibrator 5 3 7 Indication of the block temperature and set tempera...

Page 2: ...o the current state of technology This refers to measuring accuracy function as well as safe operation of the devices To grant safe operation competent and responsible behaviour of the operator is ess...

Page 3: ...and control device to be used according to EN61010 1 for the altitudes operation until 2000 m The use of the instrument for applications not expressly permitted by the above mentioned rule must adopt...

Page 4: ...essary to use exactly adapted pockets The right bore is the outer dameter of the device under test 0 5 mm 3 4 Control of temperature sensors For the test of temperature sensors a separate temperature...

Page 5: ...brass block up from ambient temperature to the adjusted set point value 3 7 Indication of the block temperature and set temperature The red 4 digit 7 segment LED PV shows continuously the temperature...

Page 6: ...ng block and test piece Before transportation or touching the heating block the block has to be cooled down 5 Recalibration The test instrument series TP18 000 E have been adjusted and checked before...

Page 7: ...a detailed description of the faults found without any fees or charge for SIKA except in the event of alternative agreements 10 Technical Data Regulation of the heating block with PID regulator Adjust...

Page 8: ...Fuse 1 6 A T 230VAC 3 15 A T 115VAC Operating temperature 0 50 C Operating humidity 30 95 RH without condensation Storage temperature 10 60 C Compliance 2014 35 EC EN 61010 2014 30 EC EN 61326 1 Dime...

Reviews: