Make shure that the operator of the equipment is instructed regularly in all applicable questions of
industrial safety and environmental protection, as well as is acquainted with the operating instructions, in
particular with the mentioned safety notes.

Temperature fuse

For safety reasons the instrument has a temperature fuse for switching off the power for the heating if the
temperature of the heating block is above 210 °C. After cooling down the heating block the unit is ready
for working as before.

The instrument has been projected and manufactured as a measuring and control device to be used
according to EN61010-1 for the altitudes operation until 2000 m.
The use of the instrument for applications not expressly permitted by the above mentioned rule must
adopt all the necessary protective measures.

The instrument 

CANNOT

 be used in 

dangerous environments (flammable or explosive)

 without

adequate protection.

The installer must ensure that EMC rules are respected, also after the instrument installation, if necessary
using proper filters.

Whenever a failure or a malfunction of the device may cause dangerous situations for persons, thing or
animals, please remember that the plant has to be equipped with additional devices which will guarantee
safety.

Attention:

 Danger of burning (block and test piece).

Before transportation or touching the heating block the block has to be cooled down.

Summary of Contents for SIKA TP 18 200 E

Page 1: ...erating position 4 3 3 Adapter sleeves 4 3 4 Control of temperature sensors 4 3 5 Front panel description 4 3 6 Switching on of the calibrator 5 3 7 Indication of the block temperature and set tempera...

Page 2: ...o the current state of technology This refers to measuring accuracy function as well as safe operation of the devices To grant safe operation competent and responsible behaviour of the operator is ess...

Page 3: ...and control device to be used according to EN61010 1 for the altitudes operation until 2000 m The use of the instrument for applications not expressly permitted by the above mentioned rule must adopt...

Page 4: ...essary to use exactly adapted pockets The right bore is the outer dameter of the device under test 0 5 mm 3 4 Control of temperature sensors For the test of temperature sensors a separate temperature...

Page 5: ...brass block up from ambient temperature to the adjusted set point value 3 7 Indication of the block temperature and set temperature The red 4 digit 7 segment LED PV shows continuously the temperature...

Page 6: ...ng block and test piece Before transportation or touching the heating block the block has to be cooled down 5 Recalibration The test instrument series TP18 000 E have been adjusted and checked before...

Page 7: ...a detailed description of the faults found without any fees or charge for SIKA except in the event of alternative agreements 10 Technical Data Regulation of the heating block with PID regulator Adjust...

Page 8: ...Fuse 1 6 A T 230VAC 3 15 A T 115VAC Operating temperature 0 50 C Operating humidity 30 95 RH without condensation Storage temperature 10 60 C Compliance 2014 35 EC EN 61010 2014 30 EC EN 61326 1 Dime...

Reviews: