3   Operation of the calibrator 

The test instrument TP 18 200 E  is a portable unit for service purposes, as well as for operating and laboratory 
tasks. 

3.1   Adjustment of the handle 

The handle can be adjusted into any required position from vertical to horizontal by pressing the lateral stop 
buttons. 

3.2   Operating position 

To guarantee an optimal temperature distribution in the heating block, the normal operating position for the 
instrument is the upright position 

3.3   Adapter sleeves 

It has to be made sure that the heat transmission resistance between the heating block and the test piece 
(thermometer, temperature sensor, temperature switch) is as low as possible.  Therefore it is necessary to use 
exactly adapted pockets. The right bore is the outer dameter of the device under test + 0.5 mm. 

3.4   Control of temperature sensors 

For the test of temperature sensors a separate temperature measuring unit is required for the connection of the 
test piece. By comparing the temperature of the external measuring unit with the block temperature you get an 
information about the status of the test piece. Please note that it takes a short time until the test piece has 
accepted the temperature of the block. 

3.5   Front panel description 

1 - Key P 

This is used to access set point of the heating block and to confirm the selection. 

2 - Key DOWN 

This is used to decrease the values of the heating block 

3 - Key UP 

This is used to increase the values of the heating block 

4 - Key U 

no function for temeprature calibrator 

5 - Led OUT1 

indicates the state of output for heating block 

6 - Led OUT2 

indicates the state of output for temperature fuse 

Summary of Contents for SIKA TP 18 200 E

Page 1: ...erating position 4 3 3 Adapter sleeves 4 3 4 Control of temperature sensors 4 3 5 Front panel description 4 3 6 Switching on of the calibrator 5 3 7 Indication of the block temperature and set tempera...

Page 2: ...o the current state of technology This refers to measuring accuracy function as well as safe operation of the devices To grant safe operation competent and responsible behaviour of the operator is ess...

Page 3: ...and control device to be used according to EN61010 1 for the altitudes operation until 2000 m The use of the instrument for applications not expressly permitted by the above mentioned rule must adopt...

Page 4: ...essary to use exactly adapted pockets The right bore is the outer dameter of the device under test 0 5 mm 3 4 Control of temperature sensors For the test of temperature sensors a separate temperature...

Page 5: ...brass block up from ambient temperature to the adjusted set point value 3 7 Indication of the block temperature and set temperature The red 4 digit 7 segment LED PV shows continuously the temperature...

Page 6: ...ng block and test piece Before transportation or touching the heating block the block has to be cooled down 5 Recalibration The test instrument series TP18 000 E have been adjusted and checked before...

Page 7: ...a detailed description of the faults found without any fees or charge for SIKA except in the event of alternative agreements 10 Technical Data Regulation of the heating block with PID regulator Adjust...

Page 8: ...Fuse 1 6 A T 230VAC 3 15 A T 115VAC Operating temperature 0 50 C Operating humidity 30 95 RH without condensation Storage temperature 10 60 C Compliance 2014 35 EC EN 61010 2014 30 EC EN 61326 1 Dime...