Time Test Facility Operation 

Press the 

ON

 button with the 

ZERO

 button held down and the timer feature initiated. (Note the instrument must be 

in the OFF state prior to this operation). In the mode the display will initially alternate between the measured 
pressure and 0.00 for time. With no pressure applied, press the 

ZERO

 button and the display will indicate ZERO and 

the pressure reading will be zeroed. The timer will remain at 0.00. 
Press the 

ZERO

 button again and the time will advance from 0.00 to a maximum 99.59 in mins and secs format. The 

display will continue to alternate between pressure and time. 
Press the 

HOLD

 button at the desired time and the display will alternate between the measured pressure and the 

paused time. Adjust the pressure to the next desired value. Press the HOLD button again and the timer will continue 
by advancing from the paused value. Press the 

HOLD

 button on completion of the test. 

CAUTION 
Product damage 

Ensure liquid media is applied to Port + only; Port - is not compatible with liquids.

Ensure liquid media contains no particulates.

Recommended that the sensor is to be positioned with Port + facing downwards; as any particulates in the
system are less likely to enter and settle within the pressure sensor.

Enure liquid media does not create a residue when drained.

Ensure liquid media is compatible with wetted materials.

Sensor is not compatible with: 

Gasoline

Motor oil

Transmission fluid

Brake & hydraulic fluids

Toluene

Cetons

Freon

Diesel fuel

Crude oil

However the use of an additional “buffer” such as a “U” tube filled with silicone oil or a large air buffer would 

make it possible to measure the above. 

Failure to comply with these instructions may result in product damage, this is not covered by the manufacturers 
warranty. 

WARRANTY 

This instrument has been tested following completion, and is warranted against faulty workmanship and materials 
for one year from the date of purchase. 
During the warranty period any defective instrument will be repaired or replaced at the discretion of the 
manufacturer.  
This warranty does not cover damage or failure resulting from misuse, accident, modification, adjustment or any 
alteration with the internal arrangement of the instrument shall absolve the manufacturer from any liability in 
respect of the instrument. 