![background image](http://html1.mh-extra.com/html/ics/pm-20/pm-20_quick-start-manual_3541050002.webp)
Time Test Facility Operation
Press the
ON
button with the
ZERO
button held down and the timer feature initiated. (Note the instrument must be
in the OFF state prior to this operation). In the mode the display will initially alternate between the measured
pressure and 0.00 for time. With no pressure applied, press the
ZERO
button and the display will indicate ZERO and
the pressure reading will be zeroed. The timer will remain at 0.00.
Press the
ZERO
button again and the time will advance from 0.00 to a maximum 99.59 in mins and secs format. The
display will continue to alternate between pressure and time.
Press the
HOLD
button at the desired time and the display will alternate between the measured pressure and the
paused time. Adjust the pressure to the next desired value. Press the HOLD button again and the timer will continue
by advancing from the paused value. Press the
HOLD
button on completion of the test.
CAUTION
Product damage
•
Ensure liquid media is applied to Port + only; Port - is not compatible with liquids.
•
Ensure liquid media contains no particulates.
•
Recommended that the sensor is to be positioned with Port + facing downwards; as any particulates in the
system are less likely to enter and settle within the pressure sensor.
•
Enure liquid media does not create a residue when drained.
•
Ensure liquid media is compatible with wetted materials.
Sensor is not compatible with:
•
Gasoline
•
Motor oil
•
Transmission fluid
•
Brake & hydraulic fluids
•
Toluene
•
Cetons
•
Freon
•
Diesel fuel
•
Crude oil
However the use of an additional “buffer” such as a “U” tube filled with silicone oil or a large air buffer would
make it possible to measure the above.
Failure to comply with these instructions may result in product damage, this is not covered by the manufacturers
warranty.
WARRANTY
This instrument has been tested following completion, and is warranted against faulty workmanship and materials
for one year from the date of purchase.
During the warranty period any defective instrument will be repaired or replaced at the discretion of the
manufacturer.
This warranty does not cover damage or failure resulting from misuse, accident, modification, adjustment or any
alteration with the internal arrangement of the instrument shall absolve the manufacturer from any liability in
respect of the instrument.