i
CR 3600
TM
– Service Manual
Guidance and Manufacture’s declaration – Electromagnetic Emis-
sions & Immunity
Table 1: Electromagnetic Emissions
Guidance and manufactures’ declaration – electromagnetic emissions
The
i
CR 3600 is intended for use in the electromagnetic environment specified
below. The customer or the user of the
i
CR 3600 should assure that it is used in
such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The
i
CR 3600 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The
i
CR 3600 is suitable for use in all
establishments, including domestic and those
directly to the public low voltage pwoer supply
network that supplies buildings used for
domestic purposes.
Harmonic
emissions IEC
61000-3-2
Class B
Voltage fluctu-
ations/flicker
emissions IEC
61000-3-3
Complies
Table 2: Electromagnetic Immunity
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment –
Guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
±
(2, 4, 6) kV
Contact
±
(2, 4, 8) kV
air
±
(2, 4, 6) kV
Contact
±
(2, 4, 8) kV
air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical
fast tran-
sient/burst
IEC
61000-4-4
±
2 kV for
power supply
lines
±
1kV for
input/output
lines
±
2 kV for
power supply
lines
±
1kV for
input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
continued on next page. . .
c
2007-2009
Confidential and Proprietary
Property of
i
CRco, Inc.
v of 61
Document # 3600-02A Rev C