5. System Operation
5.1
Power-Up
1. Make sure the power cord is properly installed in the power module.
2. Make sure the power cord is properly installed in the UPS and the UPS is plugged
into a standard wall outlet.
3. Make sure the USB 2.0 cord is properly installed in the
i
CR 3600 and computer.
4. Power on the computer.
5. Power on the
i
CR 3600 by pressing the power switch.
5.2
Cassette Handling
5.2.1
Overview
Unlike other solutions, the
i
CR 3600 does not allow for contact to the delicate phosphor
imaging plate. The hard X-ray cassette is designed to protect the plate and allow for it to be
read by the
i
CR 3600. The delicate imaging plate is embedded within a rigid cassette in
such a way that nothing touches the plate during the scanning or handling process.
5.2.2
Weight/Force Applied to Cassette
Apply only minimal weight and/or force to the cassette (for example, do
not
have a patient
stand on the cassette for weight-bearing views). Use a weight-bearing cap when necessary.
!
WARNING
Excessive weight/force may damage the cassette and may cause the CR
unit to malfunction. Do not apply excessive weight/force to the cassette.
5.2.3
Cassette Storage
All cassettes should be stored in the cleanest, driest conditions possible. Always store
cassettes in a location where they are not likely to be knocked over or damaged. With
careful handling the cassette plate system should provide years of service as there is no
known lifetime for the phosphors used in the plate system.