2. Pre-Installation
2.1
Voltage Requirements
The
i
CR 3600incorporates an international switching power supply.
i
CRco employs CE
certified medical grade universal power supplies to allow the unit to work between 90 to
253V AC 50/60Hz. In order to run the
i
CR 3600with different power types, apply voltage
to the power input module at the end of the
i
CR 3600. For international units, a local power
cord must be used that can handle the power requirements of the unit. A 13A/125V power
cord is sufficient.
2.2
Environmental
The
i
CR 3600 should not be placed in a room with a film processor present. This will void
the warranty. The humidity and temperature limits are 20 to 80% non-condensing, and 50
to 95
◦
F (10 to 35
◦
C) operating, respectively.
Installation of the
i
CR 3600 near high magnetic fields may cause the
i
CR 3600 to malfunction.
The
i
CR 3600 should not be placed in a room with an MRI, CT, or any other equipment that
produces high magnetic fields.
The room should have good ventilation. Another factor to consider prior to installing
the
i
CR 3600 is dust and particulates in the environment. The
i
CR 3600 is designed to be
resistant to dust and particulates that may be present at the installation site.
2.3
Connectivity and Power Supply
The room needs to have wall power and
should not be used with an extension cord
. Use
at least a 1300VA (780W) UPS between the wall power and the
i
CR 3600. It is also required
to have a network connection for fast technical support. Alternatively, the user must have a
phone/fax line that can be connected to the PC in order to comply with
i
CRco warranty
terms.
c
2007-2009
Confidential and Proprietary
Property of
i
CRco, Inc.
2 of 61
Document # 3600-02A Rev C
March 22, 2010