ICP DAS, ZT-2005-C8 User Manual, Version v1.0.1 Page 37
Copyright © 2014 by ICP DAS Co., Ltd. All Rights Reserved.
5.2.19
~AAEV
Desctiption
Sets to enable or disable calibration.
Syntax
~AAEV[CHKSUM](CR)
~
Delimiter character.
AA
The address of the module to be read in hexadecimal format (01 to
FF).
E
Command to enable ot disable calibration.
V
Enable or disable.
0: Disable calibration
1: Enable calibration
Response
Valid Command
!AA[CHKSUM](CR)
Invalid Command ?AA[CHKSUM](CR)
!
Delimiter character to indicate a valid command.
?
Delimiter character to indicate an invalid command.
AA
The address of the responding module in hexadecimal format (01 to
FF)
There will be no response if the command syntax is incorrect, there is a
communication error, or there is no module with the specified address.
Examples
Command ~1BE1
Response !1B
Enabled calibration of module address 27, and returns a valid response.
※
Related Commands:$AA0
、
$AA1
、
$AA0Ci
、
$AA1Ci