6
6.
System INFO
System INFO: Display serial number (SN) of the meter. This SN
is the
only one assigned to each flow meter
.
Set zero: Press
;
reset “Zero Point” which was set by the user.
Manual Totalizer: The manual totalize
r
is a separate totalize
r
.
Press to start, and press to stop
.
It is used for
isolated
flow
measurement and calculation.
Setup Menu – Calibration
Press
to
display:
, Select 2.Calibration, and the
n
and
to display 4mA
set point.
Use and to set 4mA.
After 4mA is set, press
to move to your 20mA set point.
Once 20mA is set, press
to move to next screen.
2.
Set Zero:
Press
; reset “Zero Point” which was set by the user.
3.
Low flow Cut
: Flow rate below the low flow cutoff value.
ft
ft
Press
1.
Scale factor
Refers to the ratio between
the
“actual value”
and
“reading value”. For
example, when the measurement is 2.00, and it is indicated at 1.98
on the
instrument, the scale factor reading is 2/1.98.
This means that the best scale factor constant is 1.01.
2.
Calibrate
4
&
20mA
Press