T-CAM 80 P-SERIES/T-CAM 160 XT USER MANUAL
24
9-20 Temperature Ranges
The default temperature measurement range is -20 °C ~ 150 °C (-4 °F ~ 302 °F). There may
be other options depending on purchased package.
Change the temperature range as follows:
1.
Press the Menu button.
2. Highlight Range from the Menu.
3. Press the Select button. A sub-menu will open.
4. Highlight desired temperature range in the sub-menu.
5. Press the Select button to enable the corresponding measurement. The icon means
the point is enabled. The icon means it is disabled.
6. Press the Select button to confirm and automatically close the Menu. Press Close to exit
without changes.
THE DEVICE WILL PERFORM A NUC CORRECTION AFTER A NEW
TEMPERATURE SCALE IS SELECTED.
WHEN MEASURING HIGH TEMPERATURE OBJECTS ENSURE THE
TEMPERATURE RANGE IS SET TO THE 0 °C TO 500 °C RANGE TO
OBTAIN THE MOST ACCURATE MEASUREMENT RESULTS.
9-19 Reflective Temperature
The reflective temperature is important for radiometric temperature measurement. The
imager has temperature compensation for reflective objects. Set the reflective temperature to
get the most accurate temperature measurements. In most cases, the reflected temperature
is identical to the ambient temperature.
Change the reflected temperature as follows:
1.
Press the Menu button.
2. Highlight Emissivity from the Menu.
3. Press the Select button. A sub-menu will open.
4. Highlight T. Refl. in the sub-menu.
5. Press the Select button.
6. Press the Up or Down button to modify the Reflective Temperature.
7. Press the Select button to confirm. Press Close to exit without changes.
8. Press Close to exit the menu.