SOUND DETECT & SOUND DETECT PRO USER MANUAL
43
8-6-1 Switching Application Modes
By default, the device is set to Gas Leak Mode and automatically detects and
localizes compressed air leaks. Change the device to Partial Discharge Mode
when attempting to scan for partial discharges.
Change the application mode as follows:
1.
Using the touchscreen, tap the Display Window to open the main menu.
2.
Tap the More; then, select a mode. A window appears and the user gains
access to mode settings.
3.
Tap outside window to exit the mode settings window. The main menu
closes automatically due to inactivity.
Focus
1.
3.
2.
THE USER SHOULD ENSURE THE PROPER MODE IS CHOSEN FOR THE USE
APPLICATION. GAS LEAK MODE IS FOR FINDING COMPRESSED AIR LEAKS.
PARTIAL DISCHARGE MODE IS FOR FINDING VOLTAGE PARTIAL DISCHARGES.
CHOOSE THE CORRECT SETTINGS FOR PROPER DETECTION OF AIR AND
PARTIAL DISCHARGES.
SETTING THE DISTANCE IS IMPORTANT FOR OBTAINING ACCURATE GAS
LEAK READINGS AND SOUND DISSIPATES OVER DISTANCE.
THE DISTANCE CAN ONLY BE MEASURED IN METERS.