background image

FM PLUS P SERIES IR CAMERA USER MANUAL

9

D.

Mount the FM+ Series IR Camera to a 

tripod using the ¼-20 mount.

MAKE  SURE  TRIPODS  DO  NOT  BLOCK  THE  DIRECT  PATH  OF 

PERSON(S) TO BE IMAGED TO ENSURE THE EQUIPMENT WILL NOT 

BE MOVED OR KNOCKED DOWN. USING A DIVIDING BARRIER WILL 

HELP KEEP TRIPODS SEPARATE FROM THE PATH.

C.

Plug power cord into a 

110/120V electrical outlet.

Summary of Contents for FM 320 Plus

Page 1: ...ort Control THIS DEVICE IS INTENDED FOR ADJUNCTIVE USE WITH OTHER CLINICAL DIAGNOSTIC PROCEDURES TO MEASURE HUMAN BODY TEMPERATURE VIA NON CONTACT SKIN MEASUREMENTS VISUALIZED FROM THE HUMAN FACE NOT MEANT FOR STANDALONE CLINICAL DIAGNOSTIC PROCEDURES OR TO TREAT OR DIAGNOSE PATIENTS For more information contact us 409 861 0788 sales infraredcameras com www infraredcameras com Medical Medical Edit...

Page 2: ...ATE BUT ARE PRESENTED WITHOUT WARRANTY OF ANY KIND NOTWITHSTANDING ANY OTHER WARRANTY HEREIN ALL DOCUMENT FILES AND SOFTWARE ARE PROVIDED AS IS WITH ALL FAULTS ICI DISCLAIMS ALL WARRANTIES EXPRESSED OR IMPLIED INCLUDING WITHOUT LIMITATION THOSE OF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT OR ARISING FROM A COURSE OF DEALING USAGE OR TRADE PRACTICE THE SOFTWARE LICENSE AN...

Page 3: ...ety Information 3 Intended Use 4 Technical Specifications 5 Structure 6 Package Contents 7 Installation Instructions 8 Computer Setup 9 Cleaning and Maintenance 10 Site and System Validation Checklist 11 Troubleshooting 12 About ICI Contents 4 4 4 5 6 7 8 13 16 17 18 22 ...

Page 4: ...d with care during use storage and transportation to prevent dangerous actions such as the device being hit by external force or falling from heights During transportation and storage the original packaging box must be used Prior to start of the device make sure that the power supply is properly connected If the power supply is connected incorrectly the device may be damaged Do not place any objec...

Page 5: ...Range 20 C to 50 C 68 F to 122 F Operation Range 0 C to 50 C 32 F to 122 F Storage Range 40 C to 60 C 40 F to 140 F Humidity 5 to 95 non condensing Accuracy 0 3 C 0 54 F Pixel Operability 99 Dimensions without lens 232 mm x 145 mm x 85 mm L x W x H 0 5 mm 9 13 x 5 71 x 3 35 L x W x H 0 02 Power 12 V DC 1 A 15 Watts Weight without lens 1220 g 2 69 lbs Interface RJ 45 Ethernet Video H 264 for IR and...

Page 6: ...S IR CAMERA USER MANUAL 6 5 2 Definitions of Housing Interface 1 2 1 RS485 output reserved 2 COM port reserved 3 CVBS_OUT output reserved 4 DC 12V power supply 5 RJ 45 Internet access 3 4 5 5 Structure 5 1 Appearance ...

Page 7: ... USER MANUAL 7 6 Package Contents ENSURE ALL SYSTEM EQUIPMENT AND COMPONENT ITEMS ARE PRESENT BEFORE BEGINNING INSTALLATION Temperature Reference Ethernet Cable FM Series IR Camera Power Cord x2 Software USB Drive Power Adapter ...

Page 8: ...TE AIRFLOW FROM DOORWAYS AND AIR CONDITIONING VENTILATION SYSTEMS THE SELECTED AREA SHOULD HAVE A STABLE AMBIENT TEMPERATURE BETWEEN 20 C AND 24 C 68 F TO 75 2 F AND RELATIVE HUMIDITY RANGE FROM 10 TO 50 A B Mount the Temperature Reference to a tripod using the 20 mount Plug power cord into Temperature Reference ...

Page 9: ... tripod using the 20 mount MAKE SURE TRIPODS DO NOT BLOCK THE DIRECT PATH OF PERSON S TO BE IMAGED TO ENSURE THE EQUIPMENT WILL NOT BE MOVED OR KNOCKED DOWN USING A DIVIDING BARRIER WILL HELP KEEP TRIPODS SEPARATE FROM THE PATH C Plug power cord into a 110 120V electrical outlet ...

Page 10: ...e end of the Ethernet cable into the FM Series IR Camera Ethernet port cable Plug the other end of the Ethernet cable into a computer Ethernet port G H Plug power cord into the power adapter Plug the power adapter into FM Series IR Camera barrel connector ...

Page 11: ...OW AT LEAST 45 MINUTES FOR THE TEMPERATURE REFERENCE SOURCE TO WARM UP THIS WILL PROVIDE THE MOST ACCURATE SKIN TEMPERATURE MEASUREMENTS SOFTWARE ARRIVES INSTALLED ON COMPUTERS AND TABLETS PURCHASED FROM ICI AS COMPLETE SYSTEMS AND IT WILL LAUNCH AUTOMATICALLY ON STARTUP AFTER THE SOFTWARE LAUNCHES THE USER SHOULD ENSURE THE TARGET ZONE IS IN VIEW OF THE CAMERA AND THE TEMPERATURE REFERENCE SOURCE...

Page 12: ...med with secondary evaluation methods e g an NCIT or clinical grade contact thermometer Signage should instruct for the removal of glasses headwear and masks Read the current government guidance regarding the use of telethermographic systems for skin temperature measurements Additional information can be found by reading IEC 80601 2 59 2017 Medical electrical equipment Part 2 59 Particular require...

Page 13: ... Subnet Mask should read 255 255 255 0 If it does not click in the box and fill in the correct number 10 Click Ok 8 1 Computer Setup The FM 320 Plus FM 640 Plus comes configured with static IP addresses for both the visible and infrared cameras The visible camera resides at 192 168 1 102 and the infrared camera lies at 192 168 1 102 Connecting your computer requires settings the IPv4 settings of t...

Page 14: ...1 IR Flash Skin Temperature Monitor Software IR Flash STM Software arrives installed on computers and tablets purchased from ICI as complete systems and it will launch automatically on startup A copy of the software is provided on the Software USB Drive Documentation can be found online on the IR Flash STM web page under the Downloads section or at this address https infraredcameras com ir flash s...

Page 15: ...rnet be sure to locate the computer behind a firewall Install security updates to the computer s operating system in accordance with your organization s policies Do not use the system s computer for other uses Install software updates via instructions provided by Infrared Cameras Inc only Do not install software from unknown entities Don t connect unknown hardware devices e g USB devices external ...

Page 16: ...pe the lens surface from the middle to the edge paying attention to not crack the lens or use too much liquid If the lens is still not clean replace the cloth and repeat the wiping process 9 2 Disinfecting the Camera Surface Do not use corrosive cleaning solutions on the optical glass components It is recommended to disinfect the camera surface regularly with a non corrosive sanitizing product Fol...

Page 17: ...nlight Warm or cold conductive airflow HVAC vents intakes Any radiant energy from electrical sources Direct or indirect lighting on individuals being screened Room temperature is 20 C 24 C 68 F 75 F Relative humidity is within 10 50 10 2 The Screening Background Placement is parallel and perpendicular to screening camera At a distance of 12 feet Consists of a non IR reflective homogeneous backgrou...

Page 18: ...ement on the floor V X or a set of footprints The distance is unique to each screening system Ensure the line of sight from the temperature reference source to the camera is never blocked by an individual entering or leaving the point of measurement The secondary screening area should be setup near the screening area 10 4 Typical traffic flow illustration 10 5 IR Camera Placement Level on tripod o...

Page 19: ...pod or other stable stand in front of background and positioned for visibility within the frame of the image Temperature is set at 35 C 95 F or 37 C 98 6 F Warm up time is 15 minutes 10 7 Software Set up your software according to the instructions provided Documentation can be found online on the IR Flash STM web page under the Downloads section or at this address https infraredcameras com ir flas...

Page 20: ...e readings are incorrect or facial recognition is suboptimal Close and reopen software Check that software settings are correct Fixed temperature of 35 C 95 F or 37 C 98 6 F depending on temperature reference source Temperature reference source has crosshair over it For FM FM Plus cameras ensure that proper calibration or alignment is set In Inspection System software Make sure that facial detecti...

Page 21: ...veral minutes Ensure temperatures on back of temperature reference source match Camera is away from direct sunlight or reflective light Plain background Try IR Flash STM software If already using try a newer or different version For FM Plus cameras on IR Flash STM ensure firmware update is installed ...

Page 22: ...g Institute In addition to providing custom germanium silica and sapphire optics we also build windows for enclosures as well as custom pan and tilt units We can even provide customizable explosion proof systems Our knowledge and experience stems from years of using infrared imaging and temperature measurement instruments to provide solutions to managers engineers scientists inspectors and operato...

Reviews: