FM PLUS P SERIES IR CAMERA USER MANUAL
20
11-1 Camera(s) not showing, camera(s) lagging, or software crashing
•
Close and reopen software
•
Reconnect power & Ethernet cables
•
Ensure that camera has been powered on for at least 15 minutes
•
Verify that static IP address is correct:
◦
Default IPv4 of 192.168.1.110
◦
Default subnet mask of 255.255.255.0
•
Restart computer
•
Verify that correct software is installed
•
Uninstall & reinstall software, running as administrator
•
Verify that firewall is not blocking software
•
Try IR Flash STM software. If already using, try a newer or different version.
•
For FM Plus cameras on IR Flash STM, ensure firmware update is installed
11-3 Temperature readings are incorrect or facial recognition is suboptimal
•
Close and reopen software
•
Check that software settings are correct:
◦
Fixed temperature of 35 °C (95 °F) or 37 °C (98.6 °F), depending on
temperature reference source
◦
Temperature reference source has crosshair over it
◦
For FM & FM Plus cameras, ensure that proper calibration or alignment is set
◦
In Inspection System software:
⁃
Make sure that facial detection is on
⁃
Turn on core body compensation
⁃
Set automatic compensation to environmental temperature
•
Remove masks & glasses
•
Camera is proper distance from person
•
Camera is at proper height
•
Temperature reference source in view, perpendicular to camera, but not blocked
•
Person is looking directly at camera lens, not at an angle
•
Reconnect power to camera, temperature reference source, and restart computer.
•
Temperature reference source power switch is on
11. Troubleshooting