FM PLUS P SERIES IR CAMERA USER MANUAL
17
Site:
_______________________________
System:
___________________________
Contact Person:
__________________
Date:
______________________________
**All those tasked with the initial setup of the temperature screening equipment must be
present during this training.
***This portion of the validation requires a live video feed with the instructors.
10-1 The Screening Area
□
Ensure that this area is free from:
•
Any direct or indirect (reflected) sunlight
•
Warm or cold conductive airflow
◦
HVAC vents/intakes
•
Any radiant energy from electrical sources
•
Direct or indirect lighting on individuals being screened
□
Room temperature is: 20 °C - 24°C (68 °F - 75 °F)
□
Relative humidity is within 10% – 50%
10-2 The Screening Background
□
Placement is parallel and perpendicular to screening camera
•
At a distance of ~12 feet
□
Consists of a non-IR reflective homogeneous background
•
Minimum size is: 10’ x 8’
•
White to gray flat finish
□
The temperature reference source is in front of background (direct line of sight to the
screening camera) and is framed within the screening image
10-3 The Traffic Flow
10.
Site and System Setup Validation Checklist
Continues