Cool zone center of cryoprobe 2.4 elliptic = 14.5 mm
Lesion size: 20 mm in its largest dimension
Lesion shape: oval/elliptic
14.5
–
(20/2) = 4.5 mm
Correct placement of the cryoprobe inside the lesion requires that the tip goes
beyond the lesion by 4.5 mm.
Pre-plan your trajectory needle insertion and the needle point of entry prior to the
treatment on baseline tumor measurements.
Figure 69:
Pre-Planning trajectory needle insertion in breast applications
5.2.7.2
Cryoprobe Placement Calculation during a procedure
After the system test is successfully completed, you can use the Cryoprobe Placement
Calculation screen to calculate the cryoprobe placement.
Figure 70:
Select Organ
1.
Select the organ for treatment. Once the organ is selected, the right pane of the screen will
display the Lesion size field and a keyboard.
2.
If you wish to proceed without placement calculation, press
Next
. For cryoprobe placement