Mechanical handling of hose and cryohandle
Warning
Do not try to disengage a cryoprobe before instructed since the system is
under pressure and you can be hurt by the sharp cryoprobe ejected under
pressure.
Warning
The hose and the cryoprobe still contain liquid nitrogen that can be spilled
out if you disengage the cryoprobe before you get proper system instructions
to disengage the cryoprobe.
Caution
Never use excessive force to insert or remove a cryoprobe from the
cryohandle.
If moderate force is not sufficient, contact
IceCure™
Medical for advice.
Caution
Do not pull the system by the cryohandle or hose.
The hose should be bent gently during procedure.
Emergencies and errors
Emergency Stop button
Caution
Use the Emergency Stop button only when there is no other choice.
Whenever possible, use
Extraction
to release the cryoprobe from the tissue
and use the standard shutdown procedure.
Excessive use of the Emergency Stop button may damage the system.
Warning
If a major electrical risk is detected during the cryotherapy procedure, use
the Emergency Stop button immediately.
Emergencies causing procedure halt
Warning
When a procedure halts due to an error, switch OFF the system. Call
IceCure™
Medical and describe the error shown on the screen as precisely as
possible.
Do NOT attempt to reuse the system before contacting
IceCure™
Medical.