![IceCure ProSense DMS-7063 User Manual Download Page 76](http://html1.mh-extra.com/html/icecure/prosense-dms-7063/prosense-dms-7063_user-manual_3378524076.webp)
IceCure™ Medical Ltd.
DMS-7063 rev. C
ProSense™
Confidential
User Manual
76
US User Manual
Figure 66: Wait for Functional test screen
Figure 67: Functional test screen
•
The system will check essential parameters.
•
A failure of any internal test will result in, an error message displayed on screen.
Please record the error number and follow system instructions until you are
requested to remove the cryoprobe safely from the cryohandle. You will then be
returned to the Main Menu screen. Contact
IceCure™
Medical technical service.
After successful internal test, the system will display the visual inspection screen.
You will be prompted to inspect the following:
o
The cryoprobe tip
–
make sure a small ice ball forms
o
The cryoprobe shaft
–
check that there is no ice on it
o
The saline
–
check that there are no bubbles
•
Illustrations of potential functional failures will appear on screen. Please refer to
these illustrations while inspecting the cryoprobe.